The function of 100% O2 suction key in Bennett 7200 microprocessor ventilator

M. Y. Bien, S. M. Liao, J. H. Wang

研究成果: 雜誌貢獻文章

摘要

Tracheal suctioning of a patient on a ventilator may induce hypoxemia. Bennett 7200 microprocessor ventilator has an "100% O2 suction" key which is claimed to provide 100% oxygen for two minutes and switch back automatically when this function is set. Four 7200 ventilators were tested to verify this function. We found that the mean lag time varied from 20.1 seconds to 103.3 seconds and the mean recovery time varied from 33.0 seconds to 134.6 seconds according to the difference of initial tidal volume, ventilator rate and fractional inspired oxygen setting. When the ventilator was on "100% O2 suction", the delivered tidal volume was less than the preset tidal volume. In order to hyper-oxygenate and hyper-inflate the patient to prevent suction-induced hypoxemia, those finding should be taken into consideration when the clinicians apply this function in clinical situation.

原文英語
頁(從 - 到)39-44
頁數6
期刊Chinese Medical Journal (Taipei)
45
發行號1
出版狀態已發佈 - 一月 1990
對外發佈Yes

指紋

Suction
Microcomputers
Mechanical Ventilators
Tidal Volume
Oxygen
Hypoxia

ASJC Scopus subject areas

  • Medicine(all)

引用此文

The function of 100% O2 suction key in Bennett 7200 microprocessor ventilator. / Bien, M. Y.; Liao, S. M.; Wang, J. H.

於: Chinese Medical Journal (Taipei), 卷 45, 編號 1, 01.1990, p. 39-44.

研究成果: 雜誌貢獻文章

Bien, M. Y. ; Liao, S. M. ; Wang, J. H. / The function of 100% O2 suction key in Bennett 7200 microprocessor ventilator. 於: Chinese Medical Journal (Taipei). 1990 ; 卷 45, 編號 1. 頁 39-44.
@article{50077b74c8cf45778c98cb8710890f60,
title = "The function of 100{\%} O2 suction key in Bennett 7200 microprocessor ventilator",
abstract = "Tracheal suctioning of a patient on a ventilator may induce hypoxemia. Bennett 7200 microprocessor ventilator has an {"}100{\%} O2 suction{"} key which is claimed to provide 100{\%} oxygen for two minutes and switch back automatically when this function is set. Four 7200 ventilators were tested to verify this function. We found that the mean lag time varied from 20.1 seconds to 103.3 seconds and the mean recovery time varied from 33.0 seconds to 134.6 seconds according to the difference of initial tidal volume, ventilator rate and fractional inspired oxygen setting. When the ventilator was on {"}100{\%} O2 suction{"}, the delivered tidal volume was less than the preset tidal volume. In order to hyper-oxygenate and hyper-inflate the patient to prevent suction-induced hypoxemia, those finding should be taken into consideration when the clinicians apply this function in clinical situation.",
author = "Bien, {M. Y.} and Liao, {S. M.} and Wang, {J. H.}",
year = "1990",
month = "1",
language = "English",
volume = "45",
pages = "39--44",
journal = "Journal of the Chinese Medical Association",
issn = "1726-4901",
publisher = "Elsevier Taiwan LLC",
number = "1",

}

TY - JOUR

T1 - The function of 100% O2 suction key in Bennett 7200 microprocessor ventilator

AU - Bien, M. Y.

AU - Liao, S. M.

AU - Wang, J. H.

PY - 1990/1

Y1 - 1990/1

N2 - Tracheal suctioning of a patient on a ventilator may induce hypoxemia. Bennett 7200 microprocessor ventilator has an "100% O2 suction" key which is claimed to provide 100% oxygen for two minutes and switch back automatically when this function is set. Four 7200 ventilators were tested to verify this function. We found that the mean lag time varied from 20.1 seconds to 103.3 seconds and the mean recovery time varied from 33.0 seconds to 134.6 seconds according to the difference of initial tidal volume, ventilator rate and fractional inspired oxygen setting. When the ventilator was on "100% O2 suction", the delivered tidal volume was less than the preset tidal volume. In order to hyper-oxygenate and hyper-inflate the patient to prevent suction-induced hypoxemia, those finding should be taken into consideration when the clinicians apply this function in clinical situation.

AB - Tracheal suctioning of a patient on a ventilator may induce hypoxemia. Bennett 7200 microprocessor ventilator has an "100% O2 suction" key which is claimed to provide 100% oxygen for two minutes and switch back automatically when this function is set. Four 7200 ventilators were tested to verify this function. We found that the mean lag time varied from 20.1 seconds to 103.3 seconds and the mean recovery time varied from 33.0 seconds to 134.6 seconds according to the difference of initial tidal volume, ventilator rate and fractional inspired oxygen setting. When the ventilator was on "100% O2 suction", the delivered tidal volume was less than the preset tidal volume. In order to hyper-oxygenate and hyper-inflate the patient to prevent suction-induced hypoxemia, those finding should be taken into consideration when the clinicians apply this function in clinical situation.

UR - http://www.scopus.com/inward/record.url?scp=0025152474&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0025152474&partnerID=8YFLogxK

M3 - Article

C2 - 2168238

AN - SCOPUS:0025152474

VL - 45

SP - 39

EP - 44

JO - Journal of the Chinese Medical Association

JF - Journal of the Chinese Medical Association

SN - 1726-4901

IS - 1

ER -