Is 8-mm more effective than 4-mm tip electrode catheter for ablation of typical atrial flutter?

C. F. Tsai, C. T. Tai, W. C. Yu, Y. J. Chen, M. H. Hsieh, C. E. Chiang, Y. A. Ding, M. S. Chang, S. A. Chen

研究成果: 雜誌貢獻文章同行評審

121 引文 斯高帕斯(Scopus)

摘要

Background-The prospective, randomized study comparing 4- with 8-mm tip electrodes for radiofrequency linear ablation of typical atrial flutter is not available. Methods and Results-A total of 104 consecutive patients with typical atrial flutter were randomly assigned to undergo radiofrequency linear ablation using a 4- (Group I, n=54) or 8-mm tip electrode (Group II, n=50) catheter (temperature-control model, preset 70°C). If complete bidirectional isthmus block could not be achieved after 5 pulses, the ablation catheter was changed to the other type; the maximal radiofrequency pulse number was limited to