Objective: This retrospective study was to determine factors that contribute to self-expandable metallic stent fracture in patients with tracheobronchial disease. Methods: From 2001 to 2006, 139 patients (age, 62.1 ± 15.4 years; range, 23-87 years) with benign (n = 62) and malignant (n = 77) tracheobronchial disease received 192 Ultraflex (Boston Scientific, Natick, Mass) self-expandable metallic stents (98 in patients with benign disease and 94 in patients with malignant disease). Results: Seventeen fractured self-expandable metallic stents were found; the incidence was 12.2% (17/139 patients) among patients with tracheobronchial disease. Tortuous airway (odds ratio, 4.06; 95% confidence interval, 1.04-18.34; P = .04) independently predicted self-expandable metallic stent fracture. Most self-expandable metallic stent fractures (64.7%, 11/17) were detected 500 to 1000 days after self-expandable metallic stent implantation. Clinical presentations for patients with fractured self-expandable metallic stents included dyspnea exacerbation (70.6%, 12/17) and cough (23.5%, 4/17). Conclusions: Self-expandable metallic stent fracture is not uncommon in patients with tracheobronchial disease. Tortuous airway is an independent predictor for it. Although management of the fractured self-expandable metallic stent in our study was feasible and safe, self-expandable metallic stents should be restricted to a more select population.
原文英語
頁(從 - 到)1328-1335
頁數8
期刊Journal of Thoracic and Cardiovascular Surgery
136
發行號5
DOIs
出版狀態已發佈 - 11月 1 2008
對外發佈

ASJC Scopus subject areas

  • 手術
  • 肺和呼吸系統醫學
  • 心臟病學與心血管醫學

指紋

深入研究「Factors leading to tracheobronchial self-expandable metallic stent fracture」主題。共同形成了獨特的指紋。

引用此