Early detection of implant healing process using resonance frequency analysis

Haw Ming Huang, Ching Lai Chiu, Ching-Ying Yeh, Che Tong Lin, Li Hsiang Lin, Sheng Yang Lee

研究成果: 雜誌貢獻文章

87 引文 斯高帕斯(Scopus)

摘要

In this study, in vitro and in vivo models were adopted for assessing the application of resonance frequency analysis (RFA) in the early detection of implant stability. In the in vitro tests, RF values of implants placed in bone block with predrilled cavities of 3.75 and 5.0 mm were measured and compared. Stone was used to fill the inter-space between implants and the bone blocks. Our results showed that poor initial stability conditions contributed to a lower initial RF (IRF) value (5.41±0.32 kHz) and a longer simulated healing period (41 min) than that of well-fitted conditions (9.63±0.34 kHz for IRF, 14 min for the simulated healing period, P
原文英語
頁(從 - 到)437-443
頁數7
期刊Clinical Oral Implants Research
14
發行號4
DOIs
出版狀態已發佈 - 八月 2003

    指紋

ASJC Scopus subject areas

  • Dentistry(all)

引用此