Detachment of balloon in an infant: a rare complication of balloon pulmonary valvuloplasty.

B. Hwang, B. C. Lee, J. H. Lu, C. C. Meng

研究成果: 雜誌貢獻文章同行評審

2 引文 斯高帕斯(Scopus)

摘要

A 3-month-old infant with severe tetralogy of Fallot underwent percutaneous balloon valvuloplasty. After the successful dilatation, the balloon could not be deflated completely and the partial deflated balloon was totally peeled off from the catheter at the junction of inferior vena cava and right common iliac vein when we attempted to move it out. The skin of the balloon was successfully retrieved with a snared guidewire without sequel.

原文英語
頁(從 - 到)451-454
頁數4
期刊Zhonghua Minguo xiao er ke yi xue hui za zhi [Journal]. Zhonghua Minguo xiao er ke yi xue hui
37
發行號6
出版狀態已發佈 - 1996
對外發佈

ASJC Scopus subject areas

  • 兒科、圍產兒和兒童健康

指紋

深入研究「Detachment of balloon in an infant: a rare complication of balloon pulmonary valvuloplasty.」主題。共同形成了獨特的指紋。

引用此