Design of integrated high voltage pulse generator for medical ultrasound transmitters

Deng Fong Lu, Chin Hsia, Jian Chiun Liou, Yen Chung Huang

研究成果: 雜誌貢獻文章

摘要

Design of an equivalent slew-rate monolithic pulse generator using bipolar-CMOS-DMOS (BCD) technology for medical ultrasound transmitters is presented in this paper. The pulse generator employs a floating capacitive coupling level-shifter architecture to produce a high-voltage (Vp p = 80V) output. The performance of equivalent slew-rate in the rising and falling edge is achieved by carefully choosing the value of coupling capacitors and the size of the final stage high-voltage MOSFETs of the pulse generator. The measured output pulses show the rising and falling time of 8.6 nsec and 8.5 nsec, respectively with second harmonic distortion down to −40 dBc, indicating the designed pulse generator can be used for advanced ultrasonic harmonic imaging systems.

原文英語
頁(從 - 到)1121-1127
頁數7
期刊IEICE Transactions on Communications
E102B
發行號6
DOIs
出版狀態已發佈 - 一月 1 2019

指紋

Pulse generators
Transmitters
Ultrasonics
Electric potential
Harmonic distortion
Imaging systems
Capacitors

ASJC Scopus subject areas

  • Software
  • Computer Networks and Communications
  • Electrical and Electronic Engineering

引用此文

Design of integrated high voltage pulse generator for medical ultrasound transmitters. / Lu, Deng Fong; Hsia, Chin; Liou, Jian Chiun; Huang, Yen Chung.

於: IEICE Transactions on Communications, 卷 E102B, 編號 6, 01.01.2019, p. 1121-1127.

研究成果: 雜誌貢獻文章

Lu, Deng Fong ; Hsia, Chin ; Liou, Jian Chiun ; Huang, Yen Chung. / Design of integrated high voltage pulse generator for medical ultrasound transmitters. 於: IEICE Transactions on Communications. 2019 ; 卷 E102B, 編號 6. 頁 1121-1127.
@article{63c3415b282041ab836fafee050e1960,
title = "Design of integrated high voltage pulse generator for medical ultrasound transmitters",
abstract = "Design of an equivalent slew-rate monolithic pulse generator using bipolar-CMOS-DMOS (BCD) technology for medical ultrasound transmitters is presented in this paper. The pulse generator employs a floating capacitive coupling level-shifter architecture to produce a high-voltage (Vp p = 80V) output. The performance of equivalent slew-rate in the rising and falling edge is achieved by carefully choosing the value of coupling capacitors and the size of the final stage high-voltage MOSFETs of the pulse generator. The measured output pulses show the rising and falling time of 8.6 nsec and 8.5 nsec, respectively with second harmonic distortion down to −40 dBc, indicating the designed pulse generator can be used for advanced ultrasonic harmonic imaging systems.",
keywords = "Harmonic distortion, High voltage pulse generator, Medical imaging, Ultrasound scanner",
author = "Lu, {Deng Fong} and Chin Hsia and Liou, {Jian Chiun} and Huang, {Yen Chung}",
year = "2019",
month = "1",
day = "1",
doi = "10.1587/transcom.2018HMP0001",
language = "English",
volume = "E102B",
pages = "1121--1127",
journal = "IEICE Transactions on Communications",
issn = "0916-8516",
publisher = "Maruzen Co., Ltd/Maruzen Kabushikikaisha",
number = "6",

}

TY - JOUR

T1 - Design of integrated high voltage pulse generator for medical ultrasound transmitters

AU - Lu, Deng Fong

AU - Hsia, Chin

AU - Liou, Jian Chiun

AU - Huang, Yen Chung

PY - 2019/1/1

Y1 - 2019/1/1

N2 - Design of an equivalent slew-rate monolithic pulse generator using bipolar-CMOS-DMOS (BCD) technology for medical ultrasound transmitters is presented in this paper. The pulse generator employs a floating capacitive coupling level-shifter architecture to produce a high-voltage (Vp p = 80V) output. The performance of equivalent slew-rate in the rising and falling edge is achieved by carefully choosing the value of coupling capacitors and the size of the final stage high-voltage MOSFETs of the pulse generator. The measured output pulses show the rising and falling time of 8.6 nsec and 8.5 nsec, respectively with second harmonic distortion down to −40 dBc, indicating the designed pulse generator can be used for advanced ultrasonic harmonic imaging systems.

AB - Design of an equivalent slew-rate monolithic pulse generator using bipolar-CMOS-DMOS (BCD) technology for medical ultrasound transmitters is presented in this paper. The pulse generator employs a floating capacitive coupling level-shifter architecture to produce a high-voltage (Vp p = 80V) output. The performance of equivalent slew-rate in the rising and falling edge is achieved by carefully choosing the value of coupling capacitors and the size of the final stage high-voltage MOSFETs of the pulse generator. The measured output pulses show the rising and falling time of 8.6 nsec and 8.5 nsec, respectively with second harmonic distortion down to −40 dBc, indicating the designed pulse generator can be used for advanced ultrasonic harmonic imaging systems.

KW - Harmonic distortion

KW - High voltage pulse generator

KW - Medical imaging

KW - Ultrasound scanner

UR - http://www.scopus.com/inward/record.url?scp=85072126290&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85072126290&partnerID=8YFLogxK

U2 - 10.1587/transcom.2018HMP0001

DO - 10.1587/transcom.2018HMP0001

M3 - Article

AN - SCOPUS:85072126290

VL - E102B

SP - 1121

EP - 1127

JO - IEICE Transactions on Communications

JF - IEICE Transactions on Communications

SN - 0916-8516

IS - 6

ER -