Catheter migration after implantation of an intrathecal baclofen infusion pump for severe spasticity: A case report

Tung Chou Li, Mei Hsiu Chen, Jing Shan Huang, Jun Yeen Chan, Yaun Kai Liu, Ming Hong Chen

研究成果: 雜誌貢獻文章同行評審

16 引文 斯高帕斯(Scopus)

摘要

We report a case of intrathecal baclofen infusion pump implantation complicated by migration of the catheter tip. A 55-year-old man required an intrathecal baclofen infusion for severe spasticity 4 years after a cervical spinal cord injury with incomplete tetraparesis. Twelve months after initial implantation of the device, the patient began to experience a recurrence of trunk tightness and spasticity. Subsequent X-ray and computed tomography evaluations of the catheter system revealed pooling of contrast medium outside of the intrathecal distribution in the lumbar subcutaneous region of the back and therefore migration of the pump catheter tip. At surgical revision, emphasis was placed on minimizing the length of catheter outside of the spine and securing the catheter in the supraspinous fascia with a right-angled anchor. The distance between the anchors and the entry point of the catheter into the supraspinous fascia was also reduced to prevent slipping when the patient bends forward. After surgery, the patient's spasticity improved and, 1 year later, he has experienced no further complications during follow-up, requiring an average baclofen dose of 150 μg/day. Here, we describe several surgical methods intended to secure the intrathecal catheter and prevent catheter migration. Other complications related to catheter failure are also highlighted.
原文英語
頁(從 - 到)492-497
頁數6
期刊Kaohsiung Journal of Medical Sciences
24
發行號9
DOIs
出版狀態已發佈 - 9月 2008
對外發佈

ASJC Scopus subject areas

  • 醫藥 (全部)

指紋

深入研究「Catheter migration after implantation of an intrathecal baclofen infusion pump for severe spasticity: A case report」主題。共同形成了獨特的指紋。

引用此