Amplatzer septal occluder-induced arrhythmias

Tsyr Yuh Ho, Ming Chih Tsai, Pi Chang Lee, Jen Her Lu, Chi Ming Liang, Chung Chang Laura Meng, Betau Hwang

研究成果: 雜誌貢獻文章同行評審

1 引文 斯高帕斯(Scopus)

摘要

Many complications associated with the transcatheter procedure to implant the Amplatzer septal occluder (ASO) device have emerged in the past 5 years. We discuss 2 examples here. In case 1, the ASO device was 3 mm larger than the balloon-stretched diameter. In case 2, the ASO device was the same size as the balloon-stretched diameter. One week after their discharge from the hospital, a complete atrioventricular block was detected in case 1, and complex arrhythmias consisting of atrial flutter and variable degrees of atrioventricular nodal conductions were recorded in case 2. We report our treatment of these patients and their full recovery to normal sinus rhythm at 3-week (case 1) and 6-week (case 2) follow-up examinations.

原文英語
頁(從 - 到)105-108
頁數4
期刊Journal of Medical Sciences
25
發行號2
出版狀態已發佈 - 四月 2005
對外發佈

ASJC Scopus subject areas

  • 醫藥 (全部)

指紋

深入研究「Amplatzer septal occluder-induced arrhythmias」主題。共同形成了獨特的指紋。

引用此