A Phantom Study of the Spatial Precision and Accuracy of Stereotactic Localization Using Computed Tomography Imaging with the Leksell Stereotactic System

Ali R. Tafreshi, Terrance Peng, Cheng Yu, Daniel R. Kramer, Angad S. Gogia, Morgan B. Lee, Michael F. Barbaro, Rinu Sebastian, Roberto Martin Del Campo-Vera, Kuang Hsuan Chen, Spencer S. Kellis, Brian Lee

研究成果: 雜誌貢獻文章同行評審

摘要

Background: Stereotactic localization of neurosurgical targets traditionally relies on computed tomography (CT), which is considered the optimal imaging modality for geometric accuracy. However, in-depth investigations that characterize the precision and accuracy of CT images are lacking. We used a CT phantom to examine interscanner precision and interprotocol accuracy in coordinate localization. Methods: A polymethylacrylate phantom was scanned with Toshiba Aquilion 64 and GE Healthcare LightSpeed 16 CT scanners, using both helical and incremental single-slice (SS) image acquisition protocols. The X, Y, and Z coordinates of 94 points across 6 surfaces of the phantom were physically measured. The CT scan–derived coordinates were compared with the phantom coordinates and with each other to determine accuracy and precision, respectively. Results: Using the SS imaging protocol, the mean (SD) interscanner disparity in localization was 0.93 (0.39) mm, given by the average Euclidean distance between the coordinates of the 2 scanners. This discrepancy significantly varied by axis and surface, with the greatest discrepancy in the Z-axis of 0.30 mm (95% confidence interval, 0.25–0.35; P = 0.05) and on the superior surface of 1.30 mm (95% confidence interval, 1.15–1.45; P = 0.05). SS acquisition was significantly more accurate than the helical protocol. Conclusions: We found evidence of clinically relevant inconsistency between 2 CT scanners used for stereotactic localization. SS image acquisition was superior to helical scanning with respect to localization accuracy. Interscanner consistency cannot be assumed. Institutions would benefit from identifying the errors inherent in their CT scanners.
原文英語
頁(從 - 到)e297-e307
期刊World Neurosurgery
139
DOIs
出版狀態已發佈 - 7月 2020
對外發佈

ASJC Scopus subject areas

  • 手術
  • 神經病學(臨床)

指紋

深入研究「A Phantom Study of the Spatial Precision and Accuracy of Stereotactic Localization Using Computed Tomography Imaging with the Leksell Stereotactic System」主題。共同形成了獨特的指紋。

引用此