Using 1 MHz pulse-echo ultrasound externally applied to detect mastoid effusion: Cadaver experiments

Chin Kuo Chen, Yung Liang Wan, Po Hsiang Tsui, Wen Ta Chiu, Fang Jui

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

The objective of this study is to explore the feasibility of using ultrasound to detect mastoid effusion (ME). In the past, ultrasound has been used to measure middle ear effusion (MEE) by injecting water into the external ear canal to measure echoes from the tympanic membrane, which is uncomfortable for the patient. It has been shown that air cells in the mastoid of patients with MEE are filled with fluid, which implies that ME could be a useful indicator of MEE. This study suggests using ultrasound to detect ME as a potentially noninvasive approach for MEE detection. In vitro experiments were performed on ten cadaver samples of the human ear. A single-element 1 MHz transducer was used to measure the mastoid of each cadaver before and after injecting water into the mastoid. The experimental results showed that the relative amplitudes of ultrasonic signals differed significantly between before (0.24 ± 0.09, mean ± standard deviation) and after (0.15 ± 0.03) the water injection (p <0.05, t-test), demonstrating that the ultrasonic reflection can be used to detect ME. The location of the human mastoid under the skin behind the ear allows external measurements, and hence ultrasound-based ME detection may be an alternative, noninvasive diagnostic approach to detecting MEE in the future, providing an examination that avoids discomfort.

Original languageEnglish
Pages (from-to)663-667
Number of pages5
JournalUltrasonics
Volume52
Issue number5
DOIs
Publication statusPublished - Jul 2012

Fingerprint

mastoids
echoes
middle ear
pulses
ear
ultrasonics
water injection
canals
water
standard deviation
transducers
examination

Keywords

  • Mastoid effusion
  • Middle ear effusion
  • Ultrasound

ASJC Scopus subject areas

  • Acoustics and Ultrasonics

Cite this

Using 1 MHz pulse-echo ultrasound externally applied to detect mastoid effusion : Cadaver experiments. / Chen, Chin Kuo; Wan, Yung Liang; Tsui, Po Hsiang; Chiu, Wen Ta; Jui, Fang.

In: Ultrasonics, Vol. 52, No. 5, 07.2012, p. 663-667.

Research output: Contribution to journalArticle

Chen, Chin Kuo ; Wan, Yung Liang ; Tsui, Po Hsiang ; Chiu, Wen Ta ; Jui, Fang. / Using 1 MHz pulse-echo ultrasound externally applied to detect mastoid effusion : Cadaver experiments. In: Ultrasonics. 2012 ; Vol. 52, No. 5. pp. 663-667.
@article{ad8cb180e30942418151ef482627c54d,
title = "Using 1 MHz pulse-echo ultrasound externally applied to detect mastoid effusion: Cadaver experiments",
abstract = "The objective of this study is to explore the feasibility of using ultrasound to detect mastoid effusion (ME). In the past, ultrasound has been used to measure middle ear effusion (MEE) by injecting water into the external ear canal to measure echoes from the tympanic membrane, which is uncomfortable for the patient. It has been shown that air cells in the mastoid of patients with MEE are filled with fluid, which implies that ME could be a useful indicator of MEE. This study suggests using ultrasound to detect ME as a potentially noninvasive approach for MEE detection. In vitro experiments were performed on ten cadaver samples of the human ear. A single-element 1 MHz transducer was used to measure the mastoid of each cadaver before and after injecting water into the mastoid. The experimental results showed that the relative amplitudes of ultrasonic signals differed significantly between before (0.24 ± 0.09, mean ± standard deviation) and after (0.15 ± 0.03) the water injection (p <0.05, t-test), demonstrating that the ultrasonic reflection can be used to detect ME. The location of the human mastoid under the skin behind the ear allows external measurements, and hence ultrasound-based ME detection may be an alternative, noninvasive diagnostic approach to detecting MEE in the future, providing an examination that avoids discomfort.",
keywords = "Mastoid effusion, Middle ear effusion, Ultrasound",
author = "Chen, {Chin Kuo} and Wan, {Yung Liang} and Tsui, {Po Hsiang} and Chiu, {Wen Ta} and Fang Jui",
year = "2012",
month = "7",
doi = "10.1016/j.ultras.2012.01.012",
language = "English",
volume = "52",
pages = "663--667",
journal = "Ultrasonics",
issn = "0041-624X",
publisher = "Elsevier",
number = "5",

}

TY - JOUR

T1 - Using 1 MHz pulse-echo ultrasound externally applied to detect mastoid effusion

T2 - Cadaver experiments

AU - Chen, Chin Kuo

AU - Wan, Yung Liang

AU - Tsui, Po Hsiang

AU - Chiu, Wen Ta

AU - Jui, Fang

PY - 2012/7

Y1 - 2012/7

N2 - The objective of this study is to explore the feasibility of using ultrasound to detect mastoid effusion (ME). In the past, ultrasound has been used to measure middle ear effusion (MEE) by injecting water into the external ear canal to measure echoes from the tympanic membrane, which is uncomfortable for the patient. It has been shown that air cells in the mastoid of patients with MEE are filled with fluid, which implies that ME could be a useful indicator of MEE. This study suggests using ultrasound to detect ME as a potentially noninvasive approach for MEE detection. In vitro experiments were performed on ten cadaver samples of the human ear. A single-element 1 MHz transducer was used to measure the mastoid of each cadaver before and after injecting water into the mastoid. The experimental results showed that the relative amplitudes of ultrasonic signals differed significantly between before (0.24 ± 0.09, mean ± standard deviation) and after (0.15 ± 0.03) the water injection (p <0.05, t-test), demonstrating that the ultrasonic reflection can be used to detect ME. The location of the human mastoid under the skin behind the ear allows external measurements, and hence ultrasound-based ME detection may be an alternative, noninvasive diagnostic approach to detecting MEE in the future, providing an examination that avoids discomfort.

AB - The objective of this study is to explore the feasibility of using ultrasound to detect mastoid effusion (ME). In the past, ultrasound has been used to measure middle ear effusion (MEE) by injecting water into the external ear canal to measure echoes from the tympanic membrane, which is uncomfortable for the patient. It has been shown that air cells in the mastoid of patients with MEE are filled with fluid, which implies that ME could be a useful indicator of MEE. This study suggests using ultrasound to detect ME as a potentially noninvasive approach for MEE detection. In vitro experiments were performed on ten cadaver samples of the human ear. A single-element 1 MHz transducer was used to measure the mastoid of each cadaver before and after injecting water into the mastoid. The experimental results showed that the relative amplitudes of ultrasonic signals differed significantly between before (0.24 ± 0.09, mean ± standard deviation) and after (0.15 ± 0.03) the water injection (p <0.05, t-test), demonstrating that the ultrasonic reflection can be used to detect ME. The location of the human mastoid under the skin behind the ear allows external measurements, and hence ultrasound-based ME detection may be an alternative, noninvasive diagnostic approach to detecting MEE in the future, providing an examination that avoids discomfort.

KW - Mastoid effusion

KW - Middle ear effusion

KW - Ultrasound

UR - http://www.scopus.com/inward/record.url?scp=84858449279&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84858449279&partnerID=8YFLogxK

U2 - 10.1016/j.ultras.2012.01.012

DO - 10.1016/j.ultras.2012.01.012

M3 - Article

C2 - 22348945

AN - SCOPUS:84858449279

VL - 52

SP - 663

EP - 667

JO - Ultrasonics

JF - Ultrasonics

SN - 0041-624X

IS - 5

ER -