The Potential Risk of Conical Implant-Abutment Connections

The Antirotational Ability of Cowell Implant System

Kuang Ta Yao, Hung Chan Kao, Cheng Kung Cheng, Hsu Wei Fang, Chang Hung Huang, Ming Lun Hsu

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

Background: Conical implant-abutment connections are popular because of good antibending performance. However, the cross section is round, and the antirotational ability is questionable because restorations in the oral cavity also have to bear torsional moments resulting from chewing patterns. Purpose: The purpose of this study was to investigate the antirotational ability of conical implant-abutment connections with and without an index. Materials and Methods: Conical connection implant system (Cowell Medi, Busan, South Korea) was selected. Two kinds of cyclic loading, a bending moment with (C) and without (L) a torsional moment, were respectively applied to two kinds of abutments, pure cone (N-Octa) and cone with an octagonal index (Octa). The number of cycles to fatigue and the failure modes was recorded. Morphologies of the abutments were examined with scanning electron microscopy. Results: Only group C(N-Octa) passed the fatigue test, whereas the other three groups failed because of different failure modes. In group L(N-Octa), all abutments generated rotation within 150 cycles. In groups C(Octa) and L(Octa), all abutments fractured but in different areas. Conclusions: In Cowell implant system (taper angle=7°), there was no antirotational ability in purely conical connections. Adding an octagonal index could provide an antirotational function but could compromise the antibending strength of the abutment.

Original languageEnglish
Pages (from-to)1208-1216
Number of pages9
JournalClinical Implant Dentistry and Related Research
Volume17
Issue number6
DOIs
Publication statusPublished - Dec 1 2015
Externally publishedYes

Fingerprint

Fatigue
Republic of Korea
Mastication
Electron Scanning Microscopy
Mouth

Keywords

  • Antirotational ability
  • Conical
  • Dental implant
  • Implant-abutment connection
  • Index

ASJC Scopus subject areas

  • Oral Surgery
  • Dentistry(all)

Cite this

The Potential Risk of Conical Implant-Abutment Connections : The Antirotational Ability of Cowell Implant System. / Yao, Kuang Ta; Kao, Hung Chan; Cheng, Cheng Kung; Fang, Hsu Wei; Huang, Chang Hung; Hsu, Ming Lun.

In: Clinical Implant Dentistry and Related Research, Vol. 17, No. 6, 01.12.2015, p. 1208-1216.

Research output: Contribution to journalArticle

Yao, Kuang Ta ; Kao, Hung Chan ; Cheng, Cheng Kung ; Fang, Hsu Wei ; Huang, Chang Hung ; Hsu, Ming Lun. / The Potential Risk of Conical Implant-Abutment Connections : The Antirotational Ability of Cowell Implant System. In: Clinical Implant Dentistry and Related Research. 2015 ; Vol. 17, No. 6. pp. 1208-1216.
@article{5e699e252d51458b88257cfc3427c6eb,
title = "The Potential Risk of Conical Implant-Abutment Connections: The Antirotational Ability of Cowell Implant System",
abstract = "Background: Conical implant-abutment connections are popular because of good antibending performance. However, the cross section is round, and the antirotational ability is questionable because restorations in the oral cavity also have to bear torsional moments resulting from chewing patterns. Purpose: The purpose of this study was to investigate the antirotational ability of conical implant-abutment connections with and without an index. Materials and Methods: Conical connection implant system (Cowell Medi, Busan, South Korea) was selected. Two kinds of cyclic loading, a bending moment with (C) and without (L) a torsional moment, were respectively applied to two kinds of abutments, pure cone (N-Octa) and cone with an octagonal index (Octa). The number of cycles to fatigue and the failure modes was recorded. Morphologies of the abutments were examined with scanning electron microscopy. Results: Only group C(N-Octa) passed the fatigue test, whereas the other three groups failed because of different failure modes. In group L(N-Octa), all abutments generated rotation within 150 cycles. In groups C(Octa) and L(Octa), all abutments fractured but in different areas. Conclusions: In Cowell implant system (taper angle=7°), there was no antirotational ability in purely conical connections. Adding an octagonal index could provide an antirotational function but could compromise the antibending strength of the abutment.",
keywords = "Antirotational ability, Conical, Dental implant, Implant-abutment connection, Index",
author = "Yao, {Kuang Ta} and Kao, {Hung Chan} and Cheng, {Cheng Kung} and Fang, {Hsu Wei} and Huang, {Chang Hung} and Hsu, {Ming Lun}",
year = "2015",
month = "12",
day = "1",
doi = "10.1111/cid.12219",
language = "English",
volume = "17",
pages = "1208--1216",
journal = "Clinical Implant Dentistry and Related Research",
issn = "1523-0899",
publisher = "Wiley-Blackwell",
number = "6",

}

TY - JOUR

T1 - The Potential Risk of Conical Implant-Abutment Connections

T2 - The Antirotational Ability of Cowell Implant System

AU - Yao, Kuang Ta

AU - Kao, Hung Chan

AU - Cheng, Cheng Kung

AU - Fang, Hsu Wei

AU - Huang, Chang Hung

AU - Hsu, Ming Lun

PY - 2015/12/1

Y1 - 2015/12/1

N2 - Background: Conical implant-abutment connections are popular because of good antibending performance. However, the cross section is round, and the antirotational ability is questionable because restorations in the oral cavity also have to bear torsional moments resulting from chewing patterns. Purpose: The purpose of this study was to investigate the antirotational ability of conical implant-abutment connections with and without an index. Materials and Methods: Conical connection implant system (Cowell Medi, Busan, South Korea) was selected. Two kinds of cyclic loading, a bending moment with (C) and without (L) a torsional moment, were respectively applied to two kinds of abutments, pure cone (N-Octa) and cone with an octagonal index (Octa). The number of cycles to fatigue and the failure modes was recorded. Morphologies of the abutments were examined with scanning electron microscopy. Results: Only group C(N-Octa) passed the fatigue test, whereas the other three groups failed because of different failure modes. In group L(N-Octa), all abutments generated rotation within 150 cycles. In groups C(Octa) and L(Octa), all abutments fractured but in different areas. Conclusions: In Cowell implant system (taper angle=7°), there was no antirotational ability in purely conical connections. Adding an octagonal index could provide an antirotational function but could compromise the antibending strength of the abutment.

AB - Background: Conical implant-abutment connections are popular because of good antibending performance. However, the cross section is round, and the antirotational ability is questionable because restorations in the oral cavity also have to bear torsional moments resulting from chewing patterns. Purpose: The purpose of this study was to investigate the antirotational ability of conical implant-abutment connections with and without an index. Materials and Methods: Conical connection implant system (Cowell Medi, Busan, South Korea) was selected. Two kinds of cyclic loading, a bending moment with (C) and without (L) a torsional moment, were respectively applied to two kinds of abutments, pure cone (N-Octa) and cone with an octagonal index (Octa). The number of cycles to fatigue and the failure modes was recorded. Morphologies of the abutments were examined with scanning electron microscopy. Results: Only group C(N-Octa) passed the fatigue test, whereas the other three groups failed because of different failure modes. In group L(N-Octa), all abutments generated rotation within 150 cycles. In groups C(Octa) and L(Octa), all abutments fractured but in different areas. Conclusions: In Cowell implant system (taper angle=7°), there was no antirotational ability in purely conical connections. Adding an octagonal index could provide an antirotational function but could compromise the antibending strength of the abutment.

KW - Antirotational ability

KW - Conical

KW - Dental implant

KW - Implant-abutment connection

KW - Index

UR - http://www.scopus.com/inward/record.url?scp=84954427878&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84954427878&partnerID=8YFLogxK

U2 - 10.1111/cid.12219

DO - 10.1111/cid.12219

M3 - Article

VL - 17

SP - 1208

EP - 1216

JO - Clinical Implant Dentistry and Related Research

JF - Clinical Implant Dentistry and Related Research

SN - 1523-0899

IS - 6

ER -