Quantification of non-water-suppressed MR spectra with correction for motion-induced signal reduction

Jyh Miin Lin, Shang Yueh Tsai, Hua Shan Liu, Hsiao Wen Chung, Robert V. Mulkern, Chou Min Cheng, Tzu Chen Yeh, Nan Kuei Chen

Research output: Contribution to journalArticle

13 Citations (Scopus)

Abstract

Intrascan subject movement in clinical MR spectroscopic examinations may result in inconsistent water suppression that distorts the metabolite signals, frame-to-frame variations in spectral phase and frequency, and consequent reductions in the signal-to-noise ratio due to destructive averaging. Frame-to-frame phase/frequency corrections, although reported to be successful in achieving constructive averaging, rely on consistent water suppression, which may be difficult in the presence of intrascan motion. In this study, motion correction using non-water-suppressed data acquisition is proposed to overcome the above difficulties. The time-domain matrix-pencil postprocessing method was used to extract water signals from the non-water-suppressed spectroscopic data, followed by phase and frequency corrections of the metabolite signals based on information obtained from the water signals. From in vivo experiments on seven healthy subjects at 3.0 T, quantification of metabolites using the unsuppressed water signal as a reference showed improved correlation with water-suppressed data acquired in the absence of motion (R2 = 0.9669; slope = 0.94). The metabolite concentrations derived using the proposed approach were in good agreement with literature values. Computer simulations under various degrees of frequency and phase variations further demonstrated robust performance of the time-domain postprocessing approach.

Original languageEnglish
Pages (from-to)1394-1403
Number of pages10
JournalMagnetic Resonance in Medicine
Volume62
Issue number6
DOIs
Publication statusPublished - Dec 2009
Externally publishedYes

Fingerprint

Water
Signal-To-Noise Ratio
Computer Simulation
Healthy Volunteers

Keywords

  • In vivo single voxel spectroscopy
  • Motion correction
  • Non-water suppressed MRS
  • Signal restoration
  • Spectral processing

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Quantification of non-water-suppressed MR spectra with correction for motion-induced signal reduction. / Lin, Jyh Miin; Tsai, Shang Yueh; Liu, Hua Shan; Chung, Hsiao Wen; Mulkern, Robert V.; Cheng, Chou Min; Yeh, Tzu Chen; Chen, Nan Kuei.

In: Magnetic Resonance in Medicine, Vol. 62, No. 6, 12.2009, p. 1394-1403.

Research output: Contribution to journalArticle

Lin, JM, Tsai, SY, Liu, HS, Chung, HW, Mulkern, RV, Cheng, CM, Yeh, TC & Chen, NK 2009, 'Quantification of non-water-suppressed MR spectra with correction for motion-induced signal reduction', Magnetic Resonance in Medicine, vol. 62, no. 6, pp. 1394-1403. https://doi.org/10.1002/mrm.22119
Lin, Jyh Miin ; Tsai, Shang Yueh ; Liu, Hua Shan ; Chung, Hsiao Wen ; Mulkern, Robert V. ; Cheng, Chou Min ; Yeh, Tzu Chen ; Chen, Nan Kuei. / Quantification of non-water-suppressed MR spectra with correction for motion-induced signal reduction. In: Magnetic Resonance in Medicine. 2009 ; Vol. 62, No. 6. pp. 1394-1403.
@article{b44550155f354dc0b7966326ee535cfa,
title = "Quantification of non-water-suppressed MR spectra with correction for motion-induced signal reduction",
abstract = "Intrascan subject movement in clinical MR spectroscopic examinations may result in inconsistent water suppression that distorts the metabolite signals, frame-to-frame variations in spectral phase and frequency, and consequent reductions in the signal-to-noise ratio due to destructive averaging. Frame-to-frame phase/frequency corrections, although reported to be successful in achieving constructive averaging, rely on consistent water suppression, which may be difficult in the presence of intrascan motion. In this study, motion correction using non-water-suppressed data acquisition is proposed to overcome the above difficulties. The time-domain matrix-pencil postprocessing method was used to extract water signals from the non-water-suppressed spectroscopic data, followed by phase and frequency corrections of the metabolite signals based on information obtained from the water signals. From in vivo experiments on seven healthy subjects at 3.0 T, quantification of metabolites using the unsuppressed water signal as a reference showed improved correlation with water-suppressed data acquired in the absence of motion (R2 = 0.9669; slope = 0.94). The metabolite concentrations derived using the proposed approach were in good agreement with literature values. Computer simulations under various degrees of frequency and phase variations further demonstrated robust performance of the time-domain postprocessing approach.",
keywords = "In vivo single voxel spectroscopy, Motion correction, Non-water suppressed MRS, Signal restoration, Spectral processing",
author = "Lin, {Jyh Miin} and Tsai, {Shang Yueh} and Liu, {Hua Shan} and Chung, {Hsiao Wen} and Mulkern, {Robert V.} and Cheng, {Chou Min} and Yeh, {Tzu Chen} and Chen, {Nan Kuei}",
year = "2009",
month = "12",
doi = "10.1002/mrm.22119",
language = "English",
volume = "62",
pages = "1394--1403",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "6",

}

TY - JOUR

T1 - Quantification of non-water-suppressed MR spectra with correction for motion-induced signal reduction

AU - Lin, Jyh Miin

AU - Tsai, Shang Yueh

AU - Liu, Hua Shan

AU - Chung, Hsiao Wen

AU - Mulkern, Robert V.

AU - Cheng, Chou Min

AU - Yeh, Tzu Chen

AU - Chen, Nan Kuei

PY - 2009/12

Y1 - 2009/12

N2 - Intrascan subject movement in clinical MR spectroscopic examinations may result in inconsistent water suppression that distorts the metabolite signals, frame-to-frame variations in spectral phase and frequency, and consequent reductions in the signal-to-noise ratio due to destructive averaging. Frame-to-frame phase/frequency corrections, although reported to be successful in achieving constructive averaging, rely on consistent water suppression, which may be difficult in the presence of intrascan motion. In this study, motion correction using non-water-suppressed data acquisition is proposed to overcome the above difficulties. The time-domain matrix-pencil postprocessing method was used to extract water signals from the non-water-suppressed spectroscopic data, followed by phase and frequency corrections of the metabolite signals based on information obtained from the water signals. From in vivo experiments on seven healthy subjects at 3.0 T, quantification of metabolites using the unsuppressed water signal as a reference showed improved correlation with water-suppressed data acquired in the absence of motion (R2 = 0.9669; slope = 0.94). The metabolite concentrations derived using the proposed approach were in good agreement with literature values. Computer simulations under various degrees of frequency and phase variations further demonstrated robust performance of the time-domain postprocessing approach.

AB - Intrascan subject movement in clinical MR spectroscopic examinations may result in inconsistent water suppression that distorts the metabolite signals, frame-to-frame variations in spectral phase and frequency, and consequent reductions in the signal-to-noise ratio due to destructive averaging. Frame-to-frame phase/frequency corrections, although reported to be successful in achieving constructive averaging, rely on consistent water suppression, which may be difficult in the presence of intrascan motion. In this study, motion correction using non-water-suppressed data acquisition is proposed to overcome the above difficulties. The time-domain matrix-pencil postprocessing method was used to extract water signals from the non-water-suppressed spectroscopic data, followed by phase and frequency corrections of the metabolite signals based on information obtained from the water signals. From in vivo experiments on seven healthy subjects at 3.0 T, quantification of metabolites using the unsuppressed water signal as a reference showed improved correlation with water-suppressed data acquired in the absence of motion (R2 = 0.9669; slope = 0.94). The metabolite concentrations derived using the proposed approach were in good agreement with literature values. Computer simulations under various degrees of frequency and phase variations further demonstrated robust performance of the time-domain postprocessing approach.

KW - In vivo single voxel spectroscopy

KW - Motion correction

KW - Non-water suppressed MRS

KW - Signal restoration

KW - Spectral processing

UR - http://www.scopus.com/inward/record.url?scp=73149085954&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=73149085954&partnerID=8YFLogxK

U2 - 10.1002/mrm.22119

DO - 10.1002/mrm.22119

M3 - Article

VL - 62

SP - 1394

EP - 1403

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194

IS - 6

ER -