Patient-driven loop control for ambulation function restoration in a non-invasive functional electrical stimulation system

W. L. Chen, S. C. Chen, C. C. Chen, C. H. Chou, Y. Y. Shih, Y. L. Chen, T. S. Kuo

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

Purpose. In this study, a patient-driven loop control in a non-invasive functional electrical stimulation (FES) system was designed to restore ambulation function of patients with stroke with their residual capabilities. Method. With this patient-driven loop control, patients use the electromyographic (EMG) signals from their voluntary controlled muscles in affected extremity to adjust stimulus parameters of the system. A special designed FES system generated electrical stimuli to excite the paralysed muscles through surface electrodes on the basis of the control command from the residual myoelectric signals. The EMG signals were also served as the trigger and the adjustment of stimulus parameters and thereby adding versatility of the FES system. One patient with hemiplegia was recruited to conduct clinical evaluation and treated by using the new closed-loop FES system. Results. The experimental results showed that hemiplegic could successfully control the system to restore their lost ambulation functions with the strategy of patient-driven loop control. It is revealed that the mean velocity, cadence, stride length, active ankle motion range and functional ambulation category have improved significantly from 0.22±0.17 m/s, 37.3±15.5 steps/min, 0.32±0.11 m, 10°, level 2 to 0.46±0.23 m/s, 58.2±19.1 steps/min, 0.73±0.22 m, 35°, level 4 respectively for the patient. A paired t-test indicated that differences in the EMG of the tibialis anterior and the gastrocnemius muscles between patients disabled (affected-side) foot and normal (unaffected-side) foot are not significant (p > 0.05) after 12 weeks of training. Conclusions. According to the experiment results, this patient-driven loop control can be beneficial for patient with hemiplegia to restore their ambulation functions such as dorsi-flexion and plantar-flexion. The control strategy of this study has the potential to be employed not only in the FES system but also in other assistive devices.

Original languageEnglish
Pages (from-to)65-71
Number of pages7
JournalDisability and Rehabilitation
Volume32
Issue number1
DOIs
Publication statusPublished - 2010

Fingerprint

Electric Stimulation
Walking
Hemiplegia
Foot
Skeletal Muscle
Self-Help Devices
Articular Range of Motion
Ankle
Electrodes
Extremities
Stroke
Muscles

Keywords

  • Ambulation function restoration
  • Electromyographics
  • Functional electrical stimulation
  • Hemiplegic patient
  • Patient-driven loop control

ASJC Scopus subject areas

  • Rehabilitation

Cite this

Patient-driven loop control for ambulation function restoration in a non-invasive functional electrical stimulation system. / Chen, W. L.; Chen, S. C.; Chen, C. C.; Chou, C. H.; Shih, Y. Y.; Chen, Y. L.; Kuo, T. S.

In: Disability and Rehabilitation, Vol. 32, No. 1, 2010, p. 65-71.

Research output: Contribution to journalArticle

Chen, W. L. ; Chen, S. C. ; Chen, C. C. ; Chou, C. H. ; Shih, Y. Y. ; Chen, Y. L. ; Kuo, T. S. / Patient-driven loop control for ambulation function restoration in a non-invasive functional electrical stimulation system. In: Disability and Rehabilitation. 2010 ; Vol. 32, No. 1. pp. 65-71.
@article{5d8a5b2c575544cb921bee8a87d2e076,
title = "Patient-driven loop control for ambulation function restoration in a non-invasive functional electrical stimulation system",
abstract = "Purpose. In this study, a patient-driven loop control in a non-invasive functional electrical stimulation (FES) system was designed to restore ambulation function of patients with stroke with their residual capabilities. Method. With this patient-driven loop control, patients use the electromyographic (EMG) signals from their voluntary controlled muscles in affected extremity to adjust stimulus parameters of the system. A special designed FES system generated electrical stimuli to excite the paralysed muscles through surface electrodes on the basis of the control command from the residual myoelectric signals. The EMG signals were also served as the trigger and the adjustment of stimulus parameters and thereby adding versatility of the FES system. One patient with hemiplegia was recruited to conduct clinical evaluation and treated by using the new closed-loop FES system. Results. The experimental results showed that hemiplegic could successfully control the system to restore their lost ambulation functions with the strategy of patient-driven loop control. It is revealed that the mean velocity, cadence, stride length, active ankle motion range and functional ambulation category have improved significantly from 0.22±0.17 m/s, 37.3±15.5 steps/min, 0.32±0.11 m, 10°, level 2 to 0.46±0.23 m/s, 58.2±19.1 steps/min, 0.73±0.22 m, 35°, level 4 respectively for the patient. A paired t-test indicated that differences in the EMG of the tibialis anterior and the gastrocnemius muscles between patients disabled (affected-side) foot and normal (unaffected-side) foot are not significant (p > 0.05) after 12 weeks of training. Conclusions. According to the experiment results, this patient-driven loop control can be beneficial for patient with hemiplegia to restore their ambulation functions such as dorsi-flexion and plantar-flexion. The control strategy of this study has the potential to be employed not only in the FES system but also in other assistive devices.",
keywords = "Ambulation function restoration, Electromyographics, Functional electrical stimulation, Hemiplegic patient, Patient-driven loop control",
author = "Chen, {W. L.} and Chen, {S. C.} and Chen, {C. C.} and Chou, {C. H.} and Shih, {Y. Y.} and Chen, {Y. L.} and Kuo, {T. S.}",
year = "2010",
doi = "10.3109/09638280903026564",
language = "English",
volume = "32",
pages = "65--71",
journal = "Disability and Rehabilitation",
issn = "0963-8288",
publisher = "Informa Healthcare",
number = "1",

}

TY - JOUR

T1 - Patient-driven loop control for ambulation function restoration in a non-invasive functional electrical stimulation system

AU - Chen, W. L.

AU - Chen, S. C.

AU - Chen, C. C.

AU - Chou, C. H.

AU - Shih, Y. Y.

AU - Chen, Y. L.

AU - Kuo, T. S.

PY - 2010

Y1 - 2010

N2 - Purpose. In this study, a patient-driven loop control in a non-invasive functional electrical stimulation (FES) system was designed to restore ambulation function of patients with stroke with their residual capabilities. Method. With this patient-driven loop control, patients use the electromyographic (EMG) signals from their voluntary controlled muscles in affected extremity to adjust stimulus parameters of the system. A special designed FES system generated electrical stimuli to excite the paralysed muscles through surface electrodes on the basis of the control command from the residual myoelectric signals. The EMG signals were also served as the trigger and the adjustment of stimulus parameters and thereby adding versatility of the FES system. One patient with hemiplegia was recruited to conduct clinical evaluation and treated by using the new closed-loop FES system. Results. The experimental results showed that hemiplegic could successfully control the system to restore their lost ambulation functions with the strategy of patient-driven loop control. It is revealed that the mean velocity, cadence, stride length, active ankle motion range and functional ambulation category have improved significantly from 0.22±0.17 m/s, 37.3±15.5 steps/min, 0.32±0.11 m, 10°, level 2 to 0.46±0.23 m/s, 58.2±19.1 steps/min, 0.73±0.22 m, 35°, level 4 respectively for the patient. A paired t-test indicated that differences in the EMG of the tibialis anterior and the gastrocnemius muscles between patients disabled (affected-side) foot and normal (unaffected-side) foot are not significant (p > 0.05) after 12 weeks of training. Conclusions. According to the experiment results, this patient-driven loop control can be beneficial for patient with hemiplegia to restore their ambulation functions such as dorsi-flexion and plantar-flexion. The control strategy of this study has the potential to be employed not only in the FES system but also in other assistive devices.

AB - Purpose. In this study, a patient-driven loop control in a non-invasive functional electrical stimulation (FES) system was designed to restore ambulation function of patients with stroke with their residual capabilities. Method. With this patient-driven loop control, patients use the electromyographic (EMG) signals from their voluntary controlled muscles in affected extremity to adjust stimulus parameters of the system. A special designed FES system generated electrical stimuli to excite the paralysed muscles through surface electrodes on the basis of the control command from the residual myoelectric signals. The EMG signals were also served as the trigger and the adjustment of stimulus parameters and thereby adding versatility of the FES system. One patient with hemiplegia was recruited to conduct clinical evaluation and treated by using the new closed-loop FES system. Results. The experimental results showed that hemiplegic could successfully control the system to restore their lost ambulation functions with the strategy of patient-driven loop control. It is revealed that the mean velocity, cadence, stride length, active ankle motion range and functional ambulation category have improved significantly from 0.22±0.17 m/s, 37.3±15.5 steps/min, 0.32±0.11 m, 10°, level 2 to 0.46±0.23 m/s, 58.2±19.1 steps/min, 0.73±0.22 m, 35°, level 4 respectively for the patient. A paired t-test indicated that differences in the EMG of the tibialis anterior and the gastrocnemius muscles between patients disabled (affected-side) foot and normal (unaffected-side) foot are not significant (p > 0.05) after 12 weeks of training. Conclusions. According to the experiment results, this patient-driven loop control can be beneficial for patient with hemiplegia to restore their ambulation functions such as dorsi-flexion and plantar-flexion. The control strategy of this study has the potential to be employed not only in the FES system but also in other assistive devices.

KW - Ambulation function restoration

KW - Electromyographics

KW - Functional electrical stimulation

KW - Hemiplegic patient

KW - Patient-driven loop control

UR - http://www.scopus.com/inward/record.url?scp=73349118441&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=73349118441&partnerID=8YFLogxK

U2 - 10.3109/09638280903026564

DO - 10.3109/09638280903026564

M3 - Article

VL - 32

SP - 65

EP - 71

JO - Disability and Rehabilitation

JF - Disability and Rehabilitation

SN - 0963-8288

IS - 1

ER -