Long-term efficacy and stability of atrial sensing in VDD pacing

Chun Yao Huang, Ta Chuan Tuan, Wen Shin Lee, Chien Ming Cheng, Shemg Jong Lin, Chi Woon Kong

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

Background: The efficacy and stability of atrial electrode sensing function is essential in maintaining atrioventricular (AV) synchrony and activity response in VDD pacing. Studies to evaluate the relationship between atrial sensing efficacy, stability, and implantation strategy are few. Hypothesis: This study sought to determine the possible predictors, among the recipients' underlying clinical conditions and the implantation strategies used, of achieving the most stable and effective atrial sensing. Methods: Eighty consecutive adults implanted with VDD pacemakers were studied retrospectively. Atrial electrodes were positioned where the optimal atrial potentials (APs) were measured during implantation. Interrogation parameters, including AV-synchrony ratio, were acquired immediately post implantation, at 1 year, and at 2 years after pacemaker implantation. Appropriate atrial sensing efficacy was defined as AV-synchrony ratio ≥ 90%. The stability of atrial sensing was evaluated by the variation ranges of measured APs. Results: The measured APs showed no statistical difference when age, gender, or comorbidity was considered. However, with pacing-lead bending angles <90°, patients may have presented with larger APs and narrower AP variation than patients with obtuse angles. There were no clinical parameters in our study that could have predicted the proper atrial sensing efficacy with an AV-synchrony ratio ≥ 90%. Conclusions: The long-term sensing efficacy and stability of VDD pacing is reliable. More sophisticated implantation strategies may further improve the atrial sensing stability and efficacy in VDD pacing.

Original languageEnglish
Pages (from-to)203-207
Number of pages5
JournalClinical Cardiology
Volume28
Issue number4
Publication statusPublished - Apr 2005
Externally publishedYes

Fingerprint

Electrodes
Comorbidity
Lead

Keywords

  • Atrial potentials
  • Atrioventricular synchrony
  • Bending angle
  • Efficacy
  • Fluoroscopy
  • Single-lead VDD pacemaker
  • Stability

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

Huang, C. Y., Tuan, T. C., Lee, W. S., Cheng, C. M., Lin, S. J., & Kong, C. W. (2005). Long-term efficacy and stability of atrial sensing in VDD pacing. Clinical Cardiology, 28(4), 203-207.

Long-term efficacy and stability of atrial sensing in VDD pacing. / Huang, Chun Yao; Tuan, Ta Chuan; Lee, Wen Shin; Cheng, Chien Ming; Lin, Shemg Jong; Kong, Chi Woon.

In: Clinical Cardiology, Vol. 28, No. 4, 04.2005, p. 203-207.

Research output: Contribution to journalArticle

Huang, CY, Tuan, TC, Lee, WS, Cheng, CM, Lin, SJ & Kong, CW 2005, 'Long-term efficacy and stability of atrial sensing in VDD pacing', Clinical Cardiology, vol. 28, no. 4, pp. 203-207.
Huang CY, Tuan TC, Lee WS, Cheng CM, Lin SJ, Kong CW. Long-term efficacy and stability of atrial sensing in VDD pacing. Clinical Cardiology. 2005 Apr;28(4):203-207.
Huang, Chun Yao ; Tuan, Ta Chuan ; Lee, Wen Shin ; Cheng, Chien Ming ; Lin, Shemg Jong ; Kong, Chi Woon. / Long-term efficacy and stability of atrial sensing in VDD pacing. In: Clinical Cardiology. 2005 ; Vol. 28, No. 4. pp. 203-207.
@article{8126f244c9754c63b61df384cafb8dc4,
title = "Long-term efficacy and stability of atrial sensing in VDD pacing",
abstract = "Background: The efficacy and stability of atrial electrode sensing function is essential in maintaining atrioventricular (AV) synchrony and activity response in VDD pacing. Studies to evaluate the relationship between atrial sensing efficacy, stability, and implantation strategy are few. Hypothesis: This study sought to determine the possible predictors, among the recipients' underlying clinical conditions and the implantation strategies used, of achieving the most stable and effective atrial sensing. Methods: Eighty consecutive adults implanted with VDD pacemakers were studied retrospectively. Atrial electrodes were positioned where the optimal atrial potentials (APs) were measured during implantation. Interrogation parameters, including AV-synchrony ratio, were acquired immediately post implantation, at 1 year, and at 2 years after pacemaker implantation. Appropriate atrial sensing efficacy was defined as AV-synchrony ratio ≥ 90{\%}. The stability of atrial sensing was evaluated by the variation ranges of measured APs. Results: The measured APs showed no statistical difference when age, gender, or comorbidity was considered. However, with pacing-lead bending angles <90°, patients may have presented with larger APs and narrower AP variation than patients with obtuse angles. There were no clinical parameters in our study that could have predicted the proper atrial sensing efficacy with an AV-synchrony ratio ≥ 90{\%}. Conclusions: The long-term sensing efficacy and stability of VDD pacing is reliable. More sophisticated implantation strategies may further improve the atrial sensing stability and efficacy in VDD pacing.",
keywords = "Atrial potentials, Atrioventricular synchrony, Bending angle, Efficacy, Fluoroscopy, Single-lead VDD pacemaker, Stability",
author = "Huang, {Chun Yao} and Tuan, {Ta Chuan} and Lee, {Wen Shin} and Cheng, {Chien Ming} and Lin, {Shemg Jong} and Kong, {Chi Woon}",
year = "2005",
month = "4",
language = "English",
volume = "28",
pages = "203--207",
journal = "Clinical Cardiology",
issn = "0160-9289",
publisher = "John Wiley and Sons Inc.",
number = "4",

}

TY - JOUR

T1 - Long-term efficacy and stability of atrial sensing in VDD pacing

AU - Huang, Chun Yao

AU - Tuan, Ta Chuan

AU - Lee, Wen Shin

AU - Cheng, Chien Ming

AU - Lin, Shemg Jong

AU - Kong, Chi Woon

PY - 2005/4

Y1 - 2005/4

N2 - Background: The efficacy and stability of atrial electrode sensing function is essential in maintaining atrioventricular (AV) synchrony and activity response in VDD pacing. Studies to evaluate the relationship between atrial sensing efficacy, stability, and implantation strategy are few. Hypothesis: This study sought to determine the possible predictors, among the recipients' underlying clinical conditions and the implantation strategies used, of achieving the most stable and effective atrial sensing. Methods: Eighty consecutive adults implanted with VDD pacemakers were studied retrospectively. Atrial electrodes were positioned where the optimal atrial potentials (APs) were measured during implantation. Interrogation parameters, including AV-synchrony ratio, were acquired immediately post implantation, at 1 year, and at 2 years after pacemaker implantation. Appropriate atrial sensing efficacy was defined as AV-synchrony ratio ≥ 90%. The stability of atrial sensing was evaluated by the variation ranges of measured APs. Results: The measured APs showed no statistical difference when age, gender, or comorbidity was considered. However, with pacing-lead bending angles <90°, patients may have presented with larger APs and narrower AP variation than patients with obtuse angles. There were no clinical parameters in our study that could have predicted the proper atrial sensing efficacy with an AV-synchrony ratio ≥ 90%. Conclusions: The long-term sensing efficacy and stability of VDD pacing is reliable. More sophisticated implantation strategies may further improve the atrial sensing stability and efficacy in VDD pacing.

AB - Background: The efficacy and stability of atrial electrode sensing function is essential in maintaining atrioventricular (AV) synchrony and activity response in VDD pacing. Studies to evaluate the relationship between atrial sensing efficacy, stability, and implantation strategy are few. Hypothesis: This study sought to determine the possible predictors, among the recipients' underlying clinical conditions and the implantation strategies used, of achieving the most stable and effective atrial sensing. Methods: Eighty consecutive adults implanted with VDD pacemakers were studied retrospectively. Atrial electrodes were positioned where the optimal atrial potentials (APs) were measured during implantation. Interrogation parameters, including AV-synchrony ratio, were acquired immediately post implantation, at 1 year, and at 2 years after pacemaker implantation. Appropriate atrial sensing efficacy was defined as AV-synchrony ratio ≥ 90%. The stability of atrial sensing was evaluated by the variation ranges of measured APs. Results: The measured APs showed no statistical difference when age, gender, or comorbidity was considered. However, with pacing-lead bending angles <90°, patients may have presented with larger APs and narrower AP variation than patients with obtuse angles. There were no clinical parameters in our study that could have predicted the proper atrial sensing efficacy with an AV-synchrony ratio ≥ 90%. Conclusions: The long-term sensing efficacy and stability of VDD pacing is reliable. More sophisticated implantation strategies may further improve the atrial sensing stability and efficacy in VDD pacing.

KW - Atrial potentials

KW - Atrioventricular synchrony

KW - Bending angle

KW - Efficacy

KW - Fluoroscopy

KW - Single-lead VDD pacemaker

KW - Stability

UR - http://www.scopus.com/inward/record.url?scp=16344389521&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=16344389521&partnerID=8YFLogxK

M3 - Article

C2 - 15869057

AN - SCOPUS:16344389521

VL - 28

SP - 203

EP - 207

JO - Clinical Cardiology

JF - Clinical Cardiology

SN - 0160-9289

IS - 4

ER -