Investigation of cyanocobalamin interferences to an electrochemical based hemoglobin test system

Ming-Song Hsieh, Tai Guang Wu, Jason Su, Wen Jing Cheng, Shu Min Hsieh, Wei Ming Chi

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

The structure of vitamin B-12 with a cobalt ion in the center is similar to the structure of heme with an iron center in the hemoglobin. The suspicion that the presence of vitamin B-12 in venous blood samples caused interference in readings of hemoglobin levels when using the BeneCheck hemoglobin test system, an electrochemical based strip and meter, was investigated in this study. Venous blood samples with 5 mg/dL of vitamin B-12 induced a positive error of 13% in the hemoglobin measurement, while 10 mg/dL of vitamin B-12 caused a positive error of 30% in the hemoglobin measurement. However, 1 mg/dL of vitamin B-12, a concentration much higher than that of venous blood concentrations in patients under B-12 treatment, did not cause an error in the hemoglobin measurement. These results suggest that the BeneCheck hemoglobin test system is suitable for clinical application for patients under B-12 treatment because the dosage is below 1 mg/dL.

Original languageEnglish
Pages (from-to)1570-1578
Number of pages9
JournalAnalytical Letters
Volume44
Issue number9
DOIs
Publication statusPublished - Jun 2011

Fingerprint

Hemoglobin
Vitamin B 12
Hemoglobins
Vitamins
Blood
Cobalt
Heme
Reading
Iron
Ions
Therapeutics

Keywords

  • Anemia
  • Biosensor
  • Cyanocobalamin
  • Electrochemical
  • Hemoglobin

ASJC Scopus subject areas

  • Analytical Chemistry
  • Biochemistry
  • Clinical Biochemistry
  • Biochemistry, medical
  • Electrochemistry
  • Spectroscopy

Cite this

Investigation of cyanocobalamin interferences to an electrochemical based hemoglobin test system. / Hsieh, Ming-Song; Wu, Tai Guang; Su, Jason; Cheng, Wen Jing; Hsieh, Shu Min; Chi, Wei Ming.

In: Analytical Letters, Vol. 44, No. 9, 06.2011, p. 1570-1578.

Research output: Contribution to journalArticle

Hsieh, Ming-Song ; Wu, Tai Guang ; Su, Jason ; Cheng, Wen Jing ; Hsieh, Shu Min ; Chi, Wei Ming. / Investigation of cyanocobalamin interferences to an electrochemical based hemoglobin test system. In: Analytical Letters. 2011 ; Vol. 44, No. 9. pp. 1570-1578.
@article{be2350b7be8d4c59ae9687eccd97f8ec,
title = "Investigation of cyanocobalamin interferences to an electrochemical based hemoglobin test system",
abstract = "The structure of vitamin B-12 with a cobalt ion in the center is similar to the structure of heme with an iron center in the hemoglobin. The suspicion that the presence of vitamin B-12 in venous blood samples caused interference in readings of hemoglobin levels when using the BeneCheck hemoglobin test system, an electrochemical based strip and meter, was investigated in this study. Venous blood samples with 5 mg/dL of vitamin B-12 induced a positive error of 13{\%} in the hemoglobin measurement, while 10 mg/dL of vitamin B-12 caused a positive error of 30{\%} in the hemoglobin measurement. However, 1 mg/dL of vitamin B-12, a concentration much higher than that of venous blood concentrations in patients under B-12 treatment, did not cause an error in the hemoglobin measurement. These results suggest that the BeneCheck hemoglobin test system is suitable for clinical application for patients under B-12 treatment because the dosage is below 1 mg/dL.",
keywords = "Anemia, Biosensor, Cyanocobalamin, Electrochemical, Hemoglobin",
author = "Ming-Song Hsieh and Wu, {Tai Guang} and Jason Su and Cheng, {Wen Jing} and Hsieh, {Shu Min} and Chi, {Wei Ming}",
year = "2011",
month = "6",
doi = "10.1080/00032719.2010.520389",
language = "English",
volume = "44",
pages = "1570--1578",
journal = "Analytical Letters",
issn = "0003-2719",
publisher = "Taylor and Francis Ltd.",
number = "9",

}

TY - JOUR

T1 - Investigation of cyanocobalamin interferences to an electrochemical based hemoglobin test system

AU - Hsieh, Ming-Song

AU - Wu, Tai Guang

AU - Su, Jason

AU - Cheng, Wen Jing

AU - Hsieh, Shu Min

AU - Chi, Wei Ming

PY - 2011/6

Y1 - 2011/6

N2 - The structure of vitamin B-12 with a cobalt ion in the center is similar to the structure of heme with an iron center in the hemoglobin. The suspicion that the presence of vitamin B-12 in venous blood samples caused interference in readings of hemoglobin levels when using the BeneCheck hemoglobin test system, an electrochemical based strip and meter, was investigated in this study. Venous blood samples with 5 mg/dL of vitamin B-12 induced a positive error of 13% in the hemoglobin measurement, while 10 mg/dL of vitamin B-12 caused a positive error of 30% in the hemoglobin measurement. However, 1 mg/dL of vitamin B-12, a concentration much higher than that of venous blood concentrations in patients under B-12 treatment, did not cause an error in the hemoglobin measurement. These results suggest that the BeneCheck hemoglobin test system is suitable for clinical application for patients under B-12 treatment because the dosage is below 1 mg/dL.

AB - The structure of vitamin B-12 with a cobalt ion in the center is similar to the structure of heme with an iron center in the hemoglobin. The suspicion that the presence of vitamin B-12 in venous blood samples caused interference in readings of hemoglobin levels when using the BeneCheck hemoglobin test system, an electrochemical based strip and meter, was investigated in this study. Venous blood samples with 5 mg/dL of vitamin B-12 induced a positive error of 13% in the hemoglobin measurement, while 10 mg/dL of vitamin B-12 caused a positive error of 30% in the hemoglobin measurement. However, 1 mg/dL of vitamin B-12, a concentration much higher than that of venous blood concentrations in patients under B-12 treatment, did not cause an error in the hemoglobin measurement. These results suggest that the BeneCheck hemoglobin test system is suitable for clinical application for patients under B-12 treatment because the dosage is below 1 mg/dL.

KW - Anemia

KW - Biosensor

KW - Cyanocobalamin

KW - Electrochemical

KW - Hemoglobin

UR - http://www.scopus.com/inward/record.url?scp=79959494411&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79959494411&partnerID=8YFLogxK

U2 - 10.1080/00032719.2010.520389

DO - 10.1080/00032719.2010.520389

M3 - Article

AN - SCOPUS:79959494411

VL - 44

SP - 1570

EP - 1578

JO - Analytical Letters

JF - Analytical Letters

SN - 0003-2719

IS - 9

ER -