Determination of caffeic acid in rabbit plasma by high-performance liquid chromatography

Yow Shieng Uang, Fang Lien Kang, Kuang Yang Hsu

Research output: Contribution to journalArticle

21 Citations (Scopus)

Abstract

A simple and sensitive high-performance liquid chromatographic method involving UV detection was developed for determination of caffeic acid in rabbit plasma. A Lichrosphere CN column (250 mm × 4 mm I.D., 5 μm) was used as the stationary phase and the mobile phase consisted of 2% acetic acid solution at a flow-rate of 1.0 ml/min. The UV absorbance was monitored at 320 nm. The plasma sample was acidified by the addition of 0.01 parts of concentrated phosphoric acid (85%) to maintain caffeic acid stability. After a simple clean-up procedure, the limit of quantitation achieved was 0.1 μg/ml, and the standard curve was found to be linear over the concentration ranges of 0.1-2.0 μg/ml and 0.1-40 μg/ml. The coefficient of variation for within- and between-run precision and accuracy was less than 10%, and the recovery was 82.3%.

Original languageEnglish
Pages (from-to)43-49
Number of pages7
JournalJournal of Chromatography B: Biomedical Sciences and Applications
Volume673
Issue number1
DOIs
Publication statusPublished - Nov 3 1995

Fingerprint

High performance liquid chromatography
Plasmas
Acetic Acid
Flow rate
Recovery
Liquids
caffeic acid
phosphoric acid

ASJC Scopus subject areas

  • Chemistry(all)

Cite this

Determination of caffeic acid in rabbit plasma by high-performance liquid chromatography. / Uang, Yow Shieng; Kang, Fang Lien; Hsu, Kuang Yang.

In: Journal of Chromatography B: Biomedical Sciences and Applications, Vol. 673, No. 1, 03.11.1995, p. 43-49.

Research output: Contribution to journalArticle

@article{09027643419946a6b2b7c94e1a2a036b,
title = "Determination of caffeic acid in rabbit plasma by high-performance liquid chromatography",
abstract = "A simple and sensitive high-performance liquid chromatographic method involving UV detection was developed for determination of caffeic acid in rabbit plasma. A Lichrosphere CN column (250 mm × 4 mm I.D., 5 μm) was used as the stationary phase and the mobile phase consisted of 2{\%} acetic acid solution at a flow-rate of 1.0 ml/min. The UV absorbance was monitored at 320 nm. The plasma sample was acidified by the addition of 0.01 parts of concentrated phosphoric acid (85{\%}) to maintain caffeic acid stability. After a simple clean-up procedure, the limit of quantitation achieved was 0.1 μg/ml, and the standard curve was found to be linear over the concentration ranges of 0.1-2.0 μg/ml and 0.1-40 μg/ml. The coefficient of variation for within- and between-run precision and accuracy was less than 10{\%}, and the recovery was 82.3{\%}.",
author = "Uang, {Yow Shieng} and Kang, {Fang Lien} and Hsu, {Kuang Yang}",
year = "1995",
month = "11",
day = "3",
doi = "10.1016/0378-4347(95)00243-C",
language = "English",
volume = "673",
pages = "43--49",
journal = "Journal of Chromatography B: Biomedical Sciences and Applications",
issn = "0378-4347",
publisher = "Elsevier BV",
number = "1",

}

TY - JOUR

T1 - Determination of caffeic acid in rabbit plasma by high-performance liquid chromatography

AU - Uang, Yow Shieng

AU - Kang, Fang Lien

AU - Hsu, Kuang Yang

PY - 1995/11/3

Y1 - 1995/11/3

N2 - A simple and sensitive high-performance liquid chromatographic method involving UV detection was developed for determination of caffeic acid in rabbit plasma. A Lichrosphere CN column (250 mm × 4 mm I.D., 5 μm) was used as the stationary phase and the mobile phase consisted of 2% acetic acid solution at a flow-rate of 1.0 ml/min. The UV absorbance was monitored at 320 nm. The plasma sample was acidified by the addition of 0.01 parts of concentrated phosphoric acid (85%) to maintain caffeic acid stability. After a simple clean-up procedure, the limit of quantitation achieved was 0.1 μg/ml, and the standard curve was found to be linear over the concentration ranges of 0.1-2.0 μg/ml and 0.1-40 μg/ml. The coefficient of variation for within- and between-run precision and accuracy was less than 10%, and the recovery was 82.3%.

AB - A simple and sensitive high-performance liquid chromatographic method involving UV detection was developed for determination of caffeic acid in rabbit plasma. A Lichrosphere CN column (250 mm × 4 mm I.D., 5 μm) was used as the stationary phase and the mobile phase consisted of 2% acetic acid solution at a flow-rate of 1.0 ml/min. The UV absorbance was monitored at 320 nm. The plasma sample was acidified by the addition of 0.01 parts of concentrated phosphoric acid (85%) to maintain caffeic acid stability. After a simple clean-up procedure, the limit of quantitation achieved was 0.1 μg/ml, and the standard curve was found to be linear over the concentration ranges of 0.1-2.0 μg/ml and 0.1-40 μg/ml. The coefficient of variation for within- and between-run precision and accuracy was less than 10%, and the recovery was 82.3%.

UR - http://www.scopus.com/inward/record.url?scp=0028786371&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0028786371&partnerID=8YFLogxK

U2 - 10.1016/0378-4347(95)00243-C

DO - 10.1016/0378-4347(95)00243-C

M3 - Article

C2 - 8925073

AN - SCOPUS:0028786371

VL - 673

SP - 43

EP - 49

JO - Journal of Chromatography B: Biomedical Sciences and Applications

JF - Journal of Chromatography B: Biomedical Sciences and Applications

SN - 0378-4347

IS - 1