Comparison of the BACTEC MGIT 960 and BACTEC 460TB systems for detection of mycobacteria in clinical specimens

T. S. Huang, C. S. Chen, S. S J Lee, W. K. Huang, Y. C. Liu

Research output: Contribution to journalArticle

27 Citations (Scopus)

Abstract

The reliability of the Mycobacteria Growth Indicator Tube (MGIT) 960 system for rapid detection of mycobacteria in clinical specimens was evaluated and compared to the radiometric method (BACTEC 460TB) and to mycobacterial culture on Lowenstein-Jensen (LJ) medium. Clinical specimens (n = 590) were tested without selection. A total of 121 (20.5%) isolates of mycobacteria were recovered; 98 (81.0%) of them were recovered with the BACTEC 460TB system, 86 (71.1%) were recovered with the BACTEC MGIT 960 system, and 55 (45.5%) were recovered with LJ medium (MGIT 960 versus BACTEC 640TB, p >0.05; MGIT 960 or BACTEC 460TB versus LJ, p

Original languageEnglish
Pages (from-to)279-283
Number of pages5
JournalAnnals of Clinical and Laboratory Science
Volume31
Issue number3
Publication statusPublished - 2001
Externally publishedYes

Fingerprint

Mycobacterium
Growth

Keywords

  • Automation
  • Method comparison
  • Microbiological culture
  • Mycobacterium tuberculosis
  • Radiometric assay

ASJC Scopus subject areas

  • Biochemistry, Genetics and Molecular Biology(all)
  • Clinical Biochemistry

Cite this

Comparison of the BACTEC MGIT 960 and BACTEC 460TB systems for detection of mycobacteria in clinical specimens. / Huang, T. S.; Chen, C. S.; Lee, S. S J; Huang, W. K.; Liu, Y. C.

In: Annals of Clinical and Laboratory Science, Vol. 31, No. 3, 2001, p. 279-283.

Research output: Contribution to journalArticle

@article{e6751f4c3aff4f9b906481aa77ff0fa7,
title = "Comparison of the BACTEC MGIT 960 and BACTEC 460TB systems for detection of mycobacteria in clinical specimens",
abstract = "The reliability of the Mycobacteria Growth Indicator Tube (MGIT) 960 system for rapid detection of mycobacteria in clinical specimens was evaluated and compared to the radiometric method (BACTEC 460TB) and to mycobacterial culture on Lowenstein-Jensen (LJ) medium. Clinical specimens (n = 590) were tested without selection. A total of 121 (20.5{\%}) isolates of mycobacteria were recovered; 98 (81.0{\%}) of them were recovered with the BACTEC 460TB system, 86 (71.1{\%}) were recovered with the BACTEC MGIT 960 system, and 55 (45.5{\%}) were recovered with LJ medium (MGIT 960 versus BACTEC 640TB, p >0.05; MGIT 960 or BACTEC 460TB versus LJ, p",
keywords = "Automation, Method comparison, Microbiological culture, Mycobacterium tuberculosis, Radiometric assay",
author = "Huang, {T. S.} and Chen, {C. S.} and Lee, {S. S J} and Huang, {W. K.} and Liu, {Y. C.}",
year = "2001",
language = "English",
volume = "31",
pages = "279--283",
journal = "Annals of Clinical and Laboratory Science",
issn = "0091-7370",
publisher = "Association of Clinical Scientists",
number = "3",

}

TY - JOUR

T1 - Comparison of the BACTEC MGIT 960 and BACTEC 460TB systems for detection of mycobacteria in clinical specimens

AU - Huang, T. S.

AU - Chen, C. S.

AU - Lee, S. S J

AU - Huang, W. K.

AU - Liu, Y. C.

PY - 2001

Y1 - 2001

N2 - The reliability of the Mycobacteria Growth Indicator Tube (MGIT) 960 system for rapid detection of mycobacteria in clinical specimens was evaluated and compared to the radiometric method (BACTEC 460TB) and to mycobacterial culture on Lowenstein-Jensen (LJ) medium. Clinical specimens (n = 590) were tested without selection. A total of 121 (20.5%) isolates of mycobacteria were recovered; 98 (81.0%) of them were recovered with the BACTEC 460TB system, 86 (71.1%) were recovered with the BACTEC MGIT 960 system, and 55 (45.5%) were recovered with LJ medium (MGIT 960 versus BACTEC 640TB, p >0.05; MGIT 960 or BACTEC 460TB versus LJ, p

AB - The reliability of the Mycobacteria Growth Indicator Tube (MGIT) 960 system for rapid detection of mycobacteria in clinical specimens was evaluated and compared to the radiometric method (BACTEC 460TB) and to mycobacterial culture on Lowenstein-Jensen (LJ) medium. Clinical specimens (n = 590) were tested without selection. A total of 121 (20.5%) isolates of mycobacteria were recovered; 98 (81.0%) of them were recovered with the BACTEC 460TB system, 86 (71.1%) were recovered with the BACTEC MGIT 960 system, and 55 (45.5%) were recovered with LJ medium (MGIT 960 versus BACTEC 640TB, p >0.05; MGIT 960 or BACTEC 460TB versus LJ, p

KW - Automation

KW - Method comparison

KW - Microbiological culture

KW - Mycobacterium tuberculosis

KW - Radiometric assay

UR - http://www.scopus.com/inward/record.url?scp=0034908996&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0034908996&partnerID=8YFLogxK

M3 - Article

C2 - 11508832

AN - SCOPUS:0034908996

VL - 31

SP - 279

EP - 283

JO - Annals of Clinical and Laboratory Science

JF - Annals of Clinical and Laboratory Science

SN - 0091-7370

IS - 3

ER -