Avoiding Tracheal Tube Care Accidents: Elongation of Tracheal Tube Securing Straps

Yen-Liang Chang, Ming Hsu Chen, Shih-Han Hung

Research output: Contribution to journalArticle

Abstract

Purpose: Tracheal tube securing straps are essential in maintaining adequate tube function. In this study, we attempted to test and analyze some commonly used tracheal tube securing straps. Methods: Five different straps were tested under stress in dry and wet conditions. Certain elongation properties under stress on tracheal tube securing straps should be noticed. Results: The dampening of the tracheal straps seems not likely to affect the elongation properties (P>0.5). The most affected strap is the elastic strap (3.96% & 3.59% elongation, wet: dry).The cotton twill strap is the least to be extended (0.42% & 0.63% elongation, wet: dry). Conclusion: Traditional cotton twill straps remain to be most reliable choice. We propose and recommend that all tracheal straps be Pre-Stretched before use. Protocols able to stabilize tracheal tube strap will be likely to improve patient safety and avoid tube mal position accidents.
Original languageEnglish
Pages (from-to)2-
JournalAustin J Otolaryngol
Publication statusPublished - 2014

Fingerprint

Patient Safety
Accidents

Keywords

  • Tracheal tube
  • Elongation
  • Strain
  • Securing strap

Cite this

Avoiding Tracheal Tube Care Accidents : Elongation of Tracheal Tube Securing Straps. / Chang, Yen-Liang; Chen, Ming Hsu; Hung, Shih-Han.

In: Austin J Otolaryngol, 2014, p. 2-.

Research output: Contribution to journalArticle

@article{224a9a2a45a34091927dc9bda6fdfa6d,
title = "Avoiding Tracheal Tube Care Accidents: Elongation of Tracheal Tube Securing Straps",
abstract = "Purpose: Tracheal tube securing straps are essential in maintaining adequate tube function. In this study, we attempted to test and analyze some commonly used tracheal tube securing straps. Methods: Five different straps were tested under stress in dry and wet conditions. Certain elongation properties under stress on tracheal tube securing straps should be noticed. Results: The dampening of the tracheal straps seems not likely to affect the elongation properties (P>0.5). The most affected strap is the elastic strap (3.96{\%} & 3.59{\%} elongation, wet: dry).The cotton twill strap is the least to be extended (0.42{\%} & 0.63{\%} elongation, wet: dry). Conclusion: Traditional cotton twill straps remain to be most reliable choice. We propose and recommend that all tracheal straps be Pre-Stretched before use. Protocols able to stabilize tracheal tube strap will be likely to improve patient safety and avoid tube mal position accidents.",
keywords = "Tracheal tube, Elongation, Strain, Securing strap",
author = "Yen-Liang Chang and Chen, {Ming Hsu} and Shih-Han Hung",
year = "2014",
language = "English",
pages = "2--",
journal = "Austin J Otolaryngol",

}

TY - JOUR

T1 - Avoiding Tracheal Tube Care Accidents

T2 - Elongation of Tracheal Tube Securing Straps

AU - Chang, Yen-Liang

AU - Chen, Ming Hsu

AU - Hung, Shih-Han

PY - 2014

Y1 - 2014

N2 - Purpose: Tracheal tube securing straps are essential in maintaining adequate tube function. In this study, we attempted to test and analyze some commonly used tracheal tube securing straps. Methods: Five different straps were tested under stress in dry and wet conditions. Certain elongation properties under stress on tracheal tube securing straps should be noticed. Results: The dampening of the tracheal straps seems not likely to affect the elongation properties (P>0.5). The most affected strap is the elastic strap (3.96% & 3.59% elongation, wet: dry).The cotton twill strap is the least to be extended (0.42% & 0.63% elongation, wet: dry). Conclusion: Traditional cotton twill straps remain to be most reliable choice. We propose and recommend that all tracheal straps be Pre-Stretched before use. Protocols able to stabilize tracheal tube strap will be likely to improve patient safety and avoid tube mal position accidents.

AB - Purpose: Tracheal tube securing straps are essential in maintaining adequate tube function. In this study, we attempted to test and analyze some commonly used tracheal tube securing straps. Methods: Five different straps were tested under stress in dry and wet conditions. Certain elongation properties under stress on tracheal tube securing straps should be noticed. Results: The dampening of the tracheal straps seems not likely to affect the elongation properties (P>0.5). The most affected strap is the elastic strap (3.96% & 3.59% elongation, wet: dry).The cotton twill strap is the least to be extended (0.42% & 0.63% elongation, wet: dry). Conclusion: Traditional cotton twill straps remain to be most reliable choice. We propose and recommend that all tracheal straps be Pre-Stretched before use. Protocols able to stabilize tracheal tube strap will be likely to improve patient safety and avoid tube mal position accidents.

KW - Tracheal tube

KW - Elongation

KW - Strain

KW - Securing strap

UR - http://austinpublishinggroup.com/otolaryngology/fulltext/ajo-v1-id1004.php

M3 - Article

SP - 2-

JO - Austin J Otolaryngol

JF - Austin J Otolaryngol

ER -