Cage loosening continues to be a problem after spinal fusion surgery. Past studies have investigated the interfacial strength of cages in situations of a single loosening direction. However, the human spine can move in different directions; therefore, cage loosening may occur in any direction. The purposes of this study were to develop a novel finite element analysis-based neurogenetic algorithm and to discover the spike design of vertebral body cages (VBCs) with excellent interfacial pullout strength by considering multiple cage loosening directions. Five design variables of VBCs were defined, and three-dimensional finite element models were created to predict the interfacial pullout strength of different VBCs. Then, both the artificial neural network and the genetic algorithm were applied to discover the optimum VBC. Finally, one optimum VBC and five VBCs selected from the Taguchi orthogonal array were tested and compared. The optimum spike design of the VBCs was successfully determined, and the interfacial pullout strength of the optimum VBC design was superior to that of other VBCs (17–77% increase in maximum pullout force). Non-oblique spikes were suggested to enhance their loosening resistances when multiple loosening directions were considered. In conclusion, the optimum spike design revealed excellent interfacial pullout performance of vertebral body cages in multiple loosening directions. The outcome of this study could help surgeons understand the interfacial pullout strength of VBCs in terms of their spike designs, and it could provide design direction to biomechanical engineers.

Original languageEnglish
Pages (from-to)22-32
Number of pages11
JournalJournal of Medical and Biological Engineering
Volume38
Issue number1
DOIs
Publication statusPublished - Feb 1 2018

Fingerprint

Bone
Experiments
Surgery
Fusion reactions
Genetic algorithms
Neural networks
Finite element method
Engineers
Optimum design

Keywords

  • Interface performance
  • Loosening
  • Optimization
  • Verification experiments
  • Vertebral body cage

ASJC Scopus subject areas

  • Biomedical Engineering

Cite this

@article{791b70f72fdd4fe091c9deefa598ce69,
title = "An Optimization Study for the Bone-Implant Interface Performance of Lumbar Vertebral Body Cages Using a Neurogenetic Algorithm and Verification Experiment",
abstract = "Cage loosening continues to be a problem after spinal fusion surgery. Past studies have investigated the interfacial strength of cages in situations of a single loosening direction. However, the human spine can move in different directions; therefore, cage loosening may occur in any direction. The purposes of this study were to develop a novel finite element analysis-based neurogenetic algorithm and to discover the spike design of vertebral body cages (VBCs) with excellent interfacial pullout strength by considering multiple cage loosening directions. Five design variables of VBCs were defined, and three-dimensional finite element models were created to predict the interfacial pullout strength of different VBCs. Then, both the artificial neural network and the genetic algorithm were applied to discover the optimum VBC. Finally, one optimum VBC and five VBCs selected from the Taguchi orthogonal array were tested and compared. The optimum spike design of the VBCs was successfully determined, and the interfacial pullout strength of the optimum VBC design was superior to that of other VBCs (17–77{\%} increase in maximum pullout force). Non-oblique spikes were suggested to enhance their loosening resistances when multiple loosening directions were considered. In conclusion, the optimum spike design revealed excellent interfacial pullout performance of vertebral body cages in multiple loosening directions. The outcome of this study could help surgeons understand the interfacial pullout strength of VBCs in terms of their spike designs, and it could provide design direction to biomechanical engineers.",
keywords = "Interface performance, Loosening, Optimization, Verification experiments, Vertebral body cage",
author = "Lee, {Chian Her} and Hsu, {Ching Chi} and Li Chaing",
year = "2018",
month = "2",
day = "1",
doi = "10.1007/s40846-017-0306-5",
language = "English",
volume = "38",
pages = "22--32",
journal = "Journal of Medical and Biological Engineering",
issn = "1609-0985",
publisher = "中華民國生物醫學工程學會",
number = "1",

}

TY - JOUR

T1 - An Optimization Study for the Bone-Implant Interface Performance of Lumbar Vertebral Body Cages Using a Neurogenetic Algorithm and Verification Experiment

AU - Lee, Chian Her

AU - Hsu, Ching Chi

AU - Chaing, Li

PY - 2018/2/1

Y1 - 2018/2/1

N2 - Cage loosening continues to be a problem after spinal fusion surgery. Past studies have investigated the interfacial strength of cages in situations of a single loosening direction. However, the human spine can move in different directions; therefore, cage loosening may occur in any direction. The purposes of this study were to develop a novel finite element analysis-based neurogenetic algorithm and to discover the spike design of vertebral body cages (VBCs) with excellent interfacial pullout strength by considering multiple cage loosening directions. Five design variables of VBCs were defined, and three-dimensional finite element models were created to predict the interfacial pullout strength of different VBCs. Then, both the artificial neural network and the genetic algorithm were applied to discover the optimum VBC. Finally, one optimum VBC and five VBCs selected from the Taguchi orthogonal array were tested and compared. The optimum spike design of the VBCs was successfully determined, and the interfacial pullout strength of the optimum VBC design was superior to that of other VBCs (17–77% increase in maximum pullout force). Non-oblique spikes were suggested to enhance their loosening resistances when multiple loosening directions were considered. In conclusion, the optimum spike design revealed excellent interfacial pullout performance of vertebral body cages in multiple loosening directions. The outcome of this study could help surgeons understand the interfacial pullout strength of VBCs in terms of their spike designs, and it could provide design direction to biomechanical engineers.

AB - Cage loosening continues to be a problem after spinal fusion surgery. Past studies have investigated the interfacial strength of cages in situations of a single loosening direction. However, the human spine can move in different directions; therefore, cage loosening may occur in any direction. The purposes of this study were to develop a novel finite element analysis-based neurogenetic algorithm and to discover the spike design of vertebral body cages (VBCs) with excellent interfacial pullout strength by considering multiple cage loosening directions. Five design variables of VBCs were defined, and three-dimensional finite element models were created to predict the interfacial pullout strength of different VBCs. Then, both the artificial neural network and the genetic algorithm were applied to discover the optimum VBC. Finally, one optimum VBC and five VBCs selected from the Taguchi orthogonal array were tested and compared. The optimum spike design of the VBCs was successfully determined, and the interfacial pullout strength of the optimum VBC design was superior to that of other VBCs (17–77% increase in maximum pullout force). Non-oblique spikes were suggested to enhance their loosening resistances when multiple loosening directions were considered. In conclusion, the optimum spike design revealed excellent interfacial pullout performance of vertebral body cages in multiple loosening directions. The outcome of this study could help surgeons understand the interfacial pullout strength of VBCs in terms of their spike designs, and it could provide design direction to biomechanical engineers.

KW - Interface performance

KW - Loosening

KW - Optimization

KW - Verification experiments

KW - Vertebral body cage

UR - http://www.scopus.com/inward/record.url?scp=85043332574&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85043332574&partnerID=8YFLogxK

U2 - 10.1007/s40846-017-0306-5

DO - 10.1007/s40846-017-0306-5

M3 - Article

AN - SCOPUS:85043332574

VL - 38

SP - 22

EP - 32

JO - Journal of Medical and Biological Engineering

JF - Journal of Medical and Biological Engineering

SN - 1609-0985

IS - 1

ER -