An evaluation of liquid-crystal thermometry as a screening device for intraoperative hyperthermia

D. E. Lees, W. Schuette, J. M. Bull, J. Whang-Peng, E. R. Atkinson, T. E. Macnamara

Research output: Contribution to journalArticle

28 Citations (Scopus)

Abstract

Disposable liquid-crystal temperature-trend indicators were evaluated under clinical conditions that simulated the development of intraoperative hyperthermia during anesthesia. Comparison was made to forehead thermistors for rapidity, accuracy, and linearity of response as well as correlation with esophageal and rectal thermistor recordings. The liquid-crystal monitors were comparable to the forehead thermistors in both rapidity and linearity of response but not in accuracy. A linear correlation existed with esophageal thermistor temperatures. Correlation with the rectal temperatures was not as exact. It is concluded that liquid-crystal thermometers may adequately serve as screening devices for intraoperative hyperthermia.

Original languageEnglish
Pages (from-to)669-674
Number of pages6
JournalAnesthesia and Analgesia
Volume57
Issue number6
Publication statusPublished - Dec 1 1978
Externally publishedYes

Fingerprint

Thermometry
Liquid Crystals
Fever
Forehead
Equipment and Supplies
Temperature
Thermometers
Malignant Hyperthermia

ASJC Scopus subject areas

  • Anesthesiology and Pain Medicine

Cite this

Lees, D. E., Schuette, W., Bull, J. M., Whang-Peng, J., Atkinson, E. R., & Macnamara, T. E. (1978). An evaluation of liquid-crystal thermometry as a screening device for intraoperative hyperthermia. Anesthesia and Analgesia, 57(6), 669-674.

An evaluation of liquid-crystal thermometry as a screening device for intraoperative hyperthermia. / Lees, D. E.; Schuette, W.; Bull, J. M.; Whang-Peng, J.; Atkinson, E. R.; Macnamara, T. E.

In: Anesthesia and Analgesia, Vol. 57, No. 6, 01.12.1978, p. 669-674.

Research output: Contribution to journalArticle

Lees, DE, Schuette, W, Bull, JM, Whang-Peng, J, Atkinson, ER & Macnamara, TE 1978, 'An evaluation of liquid-crystal thermometry as a screening device for intraoperative hyperthermia', Anesthesia and Analgesia, vol. 57, no. 6, pp. 669-674.
Lees, D. E. ; Schuette, W. ; Bull, J. M. ; Whang-Peng, J. ; Atkinson, E. R. ; Macnamara, T. E. / An evaluation of liquid-crystal thermometry as a screening device for intraoperative hyperthermia. In: Anesthesia and Analgesia. 1978 ; Vol. 57, No. 6. pp. 669-674.
@article{2914e33a40d4466a8e516a61d120b7b1,
title = "An evaluation of liquid-crystal thermometry as a screening device for intraoperative hyperthermia",
abstract = "Disposable liquid-crystal temperature-trend indicators were evaluated under clinical conditions that simulated the development of intraoperative hyperthermia during anesthesia. Comparison was made to forehead thermistors for rapidity, accuracy, and linearity of response as well as correlation with esophageal and rectal thermistor recordings. The liquid-crystal monitors were comparable to the forehead thermistors in both rapidity and linearity of response but not in accuracy. A linear correlation existed with esophageal thermistor temperatures. Correlation with the rectal temperatures was not as exact. It is concluded that liquid-crystal thermometers may adequately serve as screening devices for intraoperative hyperthermia.",
author = "Lees, {D. E.} and W. Schuette and Bull, {J. M.} and J. Whang-Peng and Atkinson, {E. R.} and Macnamara, {T. E.}",
year = "1978",
month = "12",
day = "1",
language = "English",
volume = "57",
pages = "669--674",
journal = "Anesthesia and Analgesia",
issn = "0003-2999",
publisher = "Lippincott Williams and Wilkins",
number = "6",

}

TY - JOUR

T1 - An evaluation of liquid-crystal thermometry as a screening device for intraoperative hyperthermia

AU - Lees, D. E.

AU - Schuette, W.

AU - Bull, J. M.

AU - Whang-Peng, J.

AU - Atkinson, E. R.

AU - Macnamara, T. E.

PY - 1978/12/1

Y1 - 1978/12/1

N2 - Disposable liquid-crystal temperature-trend indicators were evaluated under clinical conditions that simulated the development of intraoperative hyperthermia during anesthesia. Comparison was made to forehead thermistors for rapidity, accuracy, and linearity of response as well as correlation with esophageal and rectal thermistor recordings. The liquid-crystal monitors were comparable to the forehead thermistors in both rapidity and linearity of response but not in accuracy. A linear correlation existed with esophageal thermistor temperatures. Correlation with the rectal temperatures was not as exact. It is concluded that liquid-crystal thermometers may adequately serve as screening devices for intraoperative hyperthermia.

AB - Disposable liquid-crystal temperature-trend indicators were evaluated under clinical conditions that simulated the development of intraoperative hyperthermia during anesthesia. Comparison was made to forehead thermistors for rapidity, accuracy, and linearity of response as well as correlation with esophageal and rectal thermistor recordings. The liquid-crystal monitors were comparable to the forehead thermistors in both rapidity and linearity of response but not in accuracy. A linear correlation existed with esophageal thermistor temperatures. Correlation with the rectal temperatures was not as exact. It is concluded that liquid-crystal thermometers may adequately serve as screening devices for intraoperative hyperthermia.

UR - http://www.scopus.com/inward/record.url?scp=0018214642&partnerID=8YFLogxK