A measure of acoustic noise generated from transcranial magnetic stimulation coils

Sameer C. Dhamne, Raveena S. Kothare, Camilla Yu, Tsung Hsun Hsieh, Elana M. Anastasio, Lindsay Oberman, Alvaro Pascual-Leone, Alexander Rotenberg

Research output: Contribution to journalArticle

12 Citations (Scopus)

Abstract

The intensity of sound emanating from the discharge of magnetic coils used in repetitive transcranial magnetic stimulation (rTMS) can potentially cause acoustic trauma. Per Occupational Safety and Health Administration (OSHA) standards for safety of noise exposure, hearing protection is recommended beyond restricted levels of noise and time limits. We measured the sound pressure levels (SPLs) from four rTMS coils with the goal of assessing if the acoustic artifact levels are of sufficient amplitude to warrant protection from acoustic trauma per OSHA standards. We studied the SPLs at two frequencies (5 and 10 Hz), three machine outputs (MO) (60, 80 and 100%), and two distances from the coil (5 and 10 cm). We found that the SPLs were louder at closer proximity from the coil and directly dependent on the MO. We also found that in all studied conditions, SPLs were lower than the OSHA permissible thresholds for short (8 h) exposure.

Original languageEnglish
Pages (from-to)432-434
Number of pages3
JournalBrain Stimulation
Volume7
Issue number3
DOIs
Publication statusPublished - 2014

Fingerprint

Transcranial Magnetic Stimulation
Acoustics
Noise
United States Occupational Safety and Health Administration
Noise-Induced Hearing Loss
Pressure
Artifacts
Hearing
Safety

Keywords

  • Acoustic noise
  • Sound pressure levels
  • Transcranial magnetic stimulation

ASJC Scopus subject areas

  • Clinical Neurology
  • Neuroscience(all)
  • Biophysics
  • Medicine(all)

Cite this

Dhamne, S. C., Kothare, R. S., Yu, C., Hsieh, T. H., Anastasio, E. M., Oberman, L., ... Rotenberg, A. (2014). A measure of acoustic noise generated from transcranial magnetic stimulation coils. Brain Stimulation, 7(3), 432-434. https://doi.org/10.1016/j.brs.2014.01.056

A measure of acoustic noise generated from transcranial magnetic stimulation coils. / Dhamne, Sameer C.; Kothare, Raveena S.; Yu, Camilla; Hsieh, Tsung Hsun; Anastasio, Elana M.; Oberman, Lindsay; Pascual-Leone, Alvaro; Rotenberg, Alexander.

In: Brain Stimulation, Vol. 7, No. 3, 2014, p. 432-434.

Research output: Contribution to journalArticle

Dhamne, SC, Kothare, RS, Yu, C, Hsieh, TH, Anastasio, EM, Oberman, L, Pascual-Leone, A & Rotenberg, A 2014, 'A measure of acoustic noise generated from transcranial magnetic stimulation coils', Brain Stimulation, vol. 7, no. 3, pp. 432-434. https://doi.org/10.1016/j.brs.2014.01.056
Dhamne, Sameer C. ; Kothare, Raveena S. ; Yu, Camilla ; Hsieh, Tsung Hsun ; Anastasio, Elana M. ; Oberman, Lindsay ; Pascual-Leone, Alvaro ; Rotenberg, Alexander. / A measure of acoustic noise generated from transcranial magnetic stimulation coils. In: Brain Stimulation. 2014 ; Vol. 7, No. 3. pp. 432-434.
@article{07e93479ba1c44368aae4af5995fd15a,
title = "A measure of acoustic noise generated from transcranial magnetic stimulation coils",
abstract = "The intensity of sound emanating from the discharge of magnetic coils used in repetitive transcranial magnetic stimulation (rTMS) can potentially cause acoustic trauma. Per Occupational Safety and Health Administration (OSHA) standards for safety of noise exposure, hearing protection is recommended beyond restricted levels of noise and time limits. We measured the sound pressure levels (SPLs) from four rTMS coils with the goal of assessing if the acoustic artifact levels are of sufficient amplitude to warrant protection from acoustic trauma per OSHA standards. We studied the SPLs at two frequencies (5 and 10 Hz), three machine outputs (MO) (60, 80 and 100{\%}), and two distances from the coil (5 and 10 cm). We found that the SPLs were louder at closer proximity from the coil and directly dependent on the MO. We also found that in all studied conditions, SPLs were lower than the OSHA permissible thresholds for short (8 h) exposure.",
keywords = "Acoustic noise, Sound pressure levels, Transcranial magnetic stimulation",
author = "Dhamne, {Sameer C.} and Kothare, {Raveena S.} and Camilla Yu and Hsieh, {Tsung Hsun} and Anastasio, {Elana M.} and Lindsay Oberman and Alvaro Pascual-Leone and Alexander Rotenberg",
year = "2014",
doi = "10.1016/j.brs.2014.01.056",
language = "English",
volume = "7",
pages = "432--434",
journal = "Brain Stimulation",
issn = "1935-861X",
publisher = "Elsevier Inc.",
number = "3",

}

TY - JOUR

T1 - A measure of acoustic noise generated from transcranial magnetic stimulation coils

AU - Dhamne, Sameer C.

AU - Kothare, Raveena S.

AU - Yu, Camilla

AU - Hsieh, Tsung Hsun

AU - Anastasio, Elana M.

AU - Oberman, Lindsay

AU - Pascual-Leone, Alvaro

AU - Rotenberg, Alexander

PY - 2014

Y1 - 2014

N2 - The intensity of sound emanating from the discharge of magnetic coils used in repetitive transcranial magnetic stimulation (rTMS) can potentially cause acoustic trauma. Per Occupational Safety and Health Administration (OSHA) standards for safety of noise exposure, hearing protection is recommended beyond restricted levels of noise and time limits. We measured the sound pressure levels (SPLs) from four rTMS coils with the goal of assessing if the acoustic artifact levels are of sufficient amplitude to warrant protection from acoustic trauma per OSHA standards. We studied the SPLs at two frequencies (5 and 10 Hz), three machine outputs (MO) (60, 80 and 100%), and two distances from the coil (5 and 10 cm). We found that the SPLs were louder at closer proximity from the coil and directly dependent on the MO. We also found that in all studied conditions, SPLs were lower than the OSHA permissible thresholds for short (8 h) exposure.

AB - The intensity of sound emanating from the discharge of magnetic coils used in repetitive transcranial magnetic stimulation (rTMS) can potentially cause acoustic trauma. Per Occupational Safety and Health Administration (OSHA) standards for safety of noise exposure, hearing protection is recommended beyond restricted levels of noise and time limits. We measured the sound pressure levels (SPLs) from four rTMS coils with the goal of assessing if the acoustic artifact levels are of sufficient amplitude to warrant protection from acoustic trauma per OSHA standards. We studied the SPLs at two frequencies (5 and 10 Hz), three machine outputs (MO) (60, 80 and 100%), and two distances from the coil (5 and 10 cm). We found that the SPLs were louder at closer proximity from the coil and directly dependent on the MO. We also found that in all studied conditions, SPLs were lower than the OSHA permissible thresholds for short (8 h) exposure.

KW - Acoustic noise

KW - Sound pressure levels

KW - Transcranial magnetic stimulation

UR - http://www.scopus.com/inward/record.url?scp=84899923702&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84899923702&partnerID=8YFLogxK

U2 - 10.1016/j.brs.2014.01.056

DO - 10.1016/j.brs.2014.01.056

M3 - Article

C2 - 24582370

AN - SCOPUS:84899923702

VL - 7

SP - 432

EP - 434

JO - Brain Stimulation

JF - Brain Stimulation

SN - 1935-861X

IS - 3

ER -