A biomechanical analysis of triangulation of anterior vertebral double-screw fixation

Tsung Jen Huang, Robert Wen Wei Hsu, Ching Lung Tai, Weng Pin Chen

Research output: Contribution to journalArticle

13 Citations (Scopus)

Abstract

Objective. This study tested the hypothesis that triangulation of two anterior vertebral screws without penetration of the cortex offers more resistance to pullout than two screws placed in parallel and penetrated. Design. The pullout strength for two parallel or two triangulated anterior vertebral screws fixation, with a uni-cortical or bi-cortical purchase, were tested and compared to the strength of a single-screw fixation with a bi-cortical purchase. Four porcine spines (six months old) were used for biomechanical test and bone mineral density was measured for each specimen before testing. Background. The potential hazards from penetration by anterior vertebral cortex screws including neurovascular and organs injuries are well documented. However, bi-cortical screw penetration is widely recognized as necessary for good anterior spinal stabilization. The authors are not aware of any biomechanical study on the anterior placement of triangulated vertebral screws without penetration and its effect on the fixation strength of anterior vertebral device remains unclear. Methods. In this study five modes of screw fixations in lateral vertebral bodies were performed: Group A, triangulated screws with one screw penetration; Group B, triangulated screws without penetration; Group C, parallel penetrating screws; Group D, parallel nonpenetrating screws; and Group E, a single-screw with bi-cortical purchase. Biomechanical analysis with a material testing system machine was performed to determine the pull out strength of each configuration. Results. The results showed that the pullout strength in the various double-screw fixation modes were statistically increased as compared to that of the single-screw with bi-cortical purchase mode. There existed statistical differences (P

Original languageEnglish
JournalClinical Biomechanics
Volume18
Issue number6
DOIs
Publication statusPublished - Jul 2003
Externally publishedYes

Fingerprint

Materials Testing
Bone Density
Spine
Swine
Equipment and Supplies
Wounds and Injuries

Keywords

  • Anterior spinal device
  • Bi-cortical
  • Biomechanical test
  • Triangulation
  • Uni-cortical
  • Vertebral screw

ASJC Scopus subject areas

  • Orthopedics and Sports Medicine

Cite this

A biomechanical analysis of triangulation of anterior vertebral double-screw fixation. / Huang, Tsung Jen; Hsu, Robert Wen Wei; Tai, Ching Lung; Chen, Weng Pin.

In: Clinical Biomechanics, Vol. 18, No. 6, 07.2003.

Research output: Contribution to journalArticle

Huang, Tsung Jen ; Hsu, Robert Wen Wei ; Tai, Ching Lung ; Chen, Weng Pin. / A biomechanical analysis of triangulation of anterior vertebral double-screw fixation. In: Clinical Biomechanics. 2003 ; Vol. 18, No. 6.
@article{7dcdf8048da74d98a7a6eb9ed4734ede,
title = "A biomechanical analysis of triangulation of anterior vertebral double-screw fixation",
abstract = "Objective. This study tested the hypothesis that triangulation of two anterior vertebral screws without penetration of the cortex offers more resistance to pullout than two screws placed in parallel and penetrated. Design. The pullout strength for two parallel or two triangulated anterior vertebral screws fixation, with a uni-cortical or bi-cortical purchase, were tested and compared to the strength of a single-screw fixation with a bi-cortical purchase. Four porcine spines (six months old) were used for biomechanical test and bone mineral density was measured for each specimen before testing. Background. The potential hazards from penetration by anterior vertebral cortex screws including neurovascular and organs injuries are well documented. However, bi-cortical screw penetration is widely recognized as necessary for good anterior spinal stabilization. The authors are not aware of any biomechanical study on the anterior placement of triangulated vertebral screws without penetration and its effect on the fixation strength of anterior vertebral device remains unclear. Methods. In this study five modes of screw fixations in lateral vertebral bodies were performed: Group A, triangulated screws with one screw penetration; Group B, triangulated screws without penetration; Group C, parallel penetrating screws; Group D, parallel nonpenetrating screws; and Group E, a single-screw with bi-cortical purchase. Biomechanical analysis with a material testing system machine was performed to determine the pull out strength of each configuration. Results. The results showed that the pullout strength in the various double-screw fixation modes were statistically increased as compared to that of the single-screw with bi-cortical purchase mode. There existed statistical differences (P",
keywords = "Anterior spinal device, Bi-cortical, Biomechanical test, Triangulation, Uni-cortical, Vertebral screw",
author = "Huang, {Tsung Jen} and Hsu, {Robert Wen Wei} and Tai, {Ching Lung} and Chen, {Weng Pin}",
year = "2003",
month = "7",
doi = "10.1016/S0268-0033(03)00083-4",
language = "English",
volume = "18",
journal = "Clinical Biomechanics",
issn = "0268-0033",
publisher = "Elsevier Limited",
number = "6",

}

TY - JOUR

T1 - A biomechanical analysis of triangulation of anterior vertebral double-screw fixation

AU - Huang, Tsung Jen

AU - Hsu, Robert Wen Wei

AU - Tai, Ching Lung

AU - Chen, Weng Pin

PY - 2003/7

Y1 - 2003/7

N2 - Objective. This study tested the hypothesis that triangulation of two anterior vertebral screws without penetration of the cortex offers more resistance to pullout than two screws placed in parallel and penetrated. Design. The pullout strength for two parallel or two triangulated anterior vertebral screws fixation, with a uni-cortical or bi-cortical purchase, were tested and compared to the strength of a single-screw fixation with a bi-cortical purchase. Four porcine spines (six months old) were used for biomechanical test and bone mineral density was measured for each specimen before testing. Background. The potential hazards from penetration by anterior vertebral cortex screws including neurovascular and organs injuries are well documented. However, bi-cortical screw penetration is widely recognized as necessary for good anterior spinal stabilization. The authors are not aware of any biomechanical study on the anterior placement of triangulated vertebral screws without penetration and its effect on the fixation strength of anterior vertebral device remains unclear. Methods. In this study five modes of screw fixations in lateral vertebral bodies were performed: Group A, triangulated screws with one screw penetration; Group B, triangulated screws without penetration; Group C, parallel penetrating screws; Group D, parallel nonpenetrating screws; and Group E, a single-screw with bi-cortical purchase. Biomechanical analysis with a material testing system machine was performed to determine the pull out strength of each configuration. Results. The results showed that the pullout strength in the various double-screw fixation modes were statistically increased as compared to that of the single-screw with bi-cortical purchase mode. There existed statistical differences (P

AB - Objective. This study tested the hypothesis that triangulation of two anterior vertebral screws without penetration of the cortex offers more resistance to pullout than two screws placed in parallel and penetrated. Design. The pullout strength for two parallel or two triangulated anterior vertebral screws fixation, with a uni-cortical or bi-cortical purchase, were tested and compared to the strength of a single-screw fixation with a bi-cortical purchase. Four porcine spines (six months old) were used for biomechanical test and bone mineral density was measured for each specimen before testing. Background. The potential hazards from penetration by anterior vertebral cortex screws including neurovascular and organs injuries are well documented. However, bi-cortical screw penetration is widely recognized as necessary for good anterior spinal stabilization. The authors are not aware of any biomechanical study on the anterior placement of triangulated vertebral screws without penetration and its effect on the fixation strength of anterior vertebral device remains unclear. Methods. In this study five modes of screw fixations in lateral vertebral bodies were performed: Group A, triangulated screws with one screw penetration; Group B, triangulated screws without penetration; Group C, parallel penetrating screws; Group D, parallel nonpenetrating screws; and Group E, a single-screw with bi-cortical purchase. Biomechanical analysis with a material testing system machine was performed to determine the pull out strength of each configuration. Results. The results showed that the pullout strength in the various double-screw fixation modes were statistically increased as compared to that of the single-screw with bi-cortical purchase mode. There existed statistical differences (P

KW - Anterior spinal device

KW - Bi-cortical

KW - Biomechanical test

KW - Triangulation

KW - Uni-cortical

KW - Vertebral screw

UR - http://www.scopus.com/inward/record.url?scp=0038386765&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0038386765&partnerID=8YFLogxK

U2 - 10.1016/S0268-0033(03)00083-4

DO - 10.1016/S0268-0033(03)00083-4

M3 - Article

VL - 18

JO - Clinical Biomechanics

JF - Clinical Biomechanics

SN - 0268-0033

IS - 6

ER -